AngioDynamics

4 recalls Categories: Surgical Instruments

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Jan 30, 2024 Other Medical Devices Nationwide View Details →

Printed Instructions for Use (IFU) not shipped with device. The Instructions for Use contain the Warnings, Precautions, Contraindications, and Procedures for the Preparation, Implantation, and Post Operative Care of the implantable port, failure to follow the Instructions for Use may result in patient injury.

Sep 20, 2022 Implants & Prosthetics Nationwide View Details →

OPS Fiber packaged within specific VenaCure EVLT Procedure Kits does not meet the required dimensional specification, may result in damage to the sheath, including separation or the inability to move the fiber within the sheath. A damaged sheath will necessitate removal from the patient, which may result in procedural delay and/or surgical intervention

Aug 10, 2022 Other Medical Devices Nationwide View Details →

Certain AngioDynamics PICC Kits contain BD PosiFlush SF Saline Flush Syringes which were recalled by BD Medical on April 16, 2020.

May 4, 2020 Diagnostic Equipment Nationwide View Details →

Certain AngioDynamics PICC Kits contain BD PosiFlush SF Saline Flush Syringes which were recalled by BD Medical on April 16, 2020.

May 4, 2020 Other Medical Devices Nationwide View Details →

BioFlo PICC and Xcela PICC Convenience Kits - Product Usage: are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients; the sampling of blood; and for power injection of contrast media. Item Numbers H96560M01162101 H96560M0116291 H96560M0133531 H96560M0134211 H96560M0198811 H96560M0224741 H96560M0224751 H96560M0335621 H96560M0367041 H96560M0367051 H96560M0367061 H96560M05019111 H96560M0501981 H96560M1201961 H96560M1201971 H96560M12202101 H96560M12202111 H96560M12202121 H96560M1303151 H96560M1303171 H96560M1303191 H96560M13184101 H96560M13184111 H96560M1318491 H96560M1401411 H96560M1401421 H96560M1401431 H96560M1401441 H96560M1406061 H96560M1406071 H96560M1633931 H96560M1803441 H96560M1803451 H96560M18157111 H96560M18157121 H96560M1841151 H96560M1841161 H96560M1841171 H96560M1907131 H96560M1907141 H96560M2204441 H96560M2208511 H96560M2208521 H96560M2322421 H96560M2500961 H96560M2500971 H96560M610115911 H96560M610901011 H96560M611025121 H96560M611025131 H96560M611052821 H96560M611056021 H96560M611056031 H96560M611095821 H96560M611095831 H965PKM030667S1 H965PKM120451S1 H965PKM1220226S1 H965PKM1300011S1 H965PKM1303119S1 H965PKM131935S1 H965PKM150124S1 H965PKM150125S1 H965PKM150127S1 H965PKM220851S1 H965PKM220854S1 H965PKM3092662S1 H965PKM61040701S1 H965PKM61066141S1 H965PKM61095321S1 H965PKM61095324S1 Cat. No. 60M0116210 60M011629 60M013353 60M013421 60M019881 60M022474 60M022475 60M033562 60M036704 60M036705 60M036706 60M0501911 60M050198 60M120196 60M120197 60M1220210 60M1220211 60M1220212 60M130315 60M130317 60M130319 60M1318410 60M1318411 60M131849 60M140141 60M140142 60M140143 60M140144 60M140606 60M140607 60M163393 60M180344 60M180345 60M1815711 60M1815712 60M184115 60M184116 60M184117 60M190713 60M190714 60M220444 60M220851 60M220852 60M232242 60M250096 60M250097 60M61011591 60M61090101 60M61102512 60M61102513 60M61105282 60M61105602 60M61105603 60M61109582 60M61109583 PKM030667S PKM120451S PKM1220226S PKM1300011S PKM1303119S PKM131935S PKM150124S PKM150125S PKM150127S PKM220851S PKM220854S PKM3092662S PKM61040701S PKM61066141S PKM61095321S PKM61095324S

Class I - Dangerous

Certain AngioDynamics PICC Kits contain BD PosiFlush SF Saline Flush Syringes which were recalled by BD Medical on April 16, 2020.

May 4, 2020 Other Medical Devices Nationwide View Details →

Certain AngioDynamics PICC Kits contain BD PosiFlush SF Saline Flush Syringes which were recalled by BD Medical on April 16, 2020.

May 4, 2020 Diagnostic Equipment Nationwide View Details →

May result in readings outside of the expected accuracy range.

Nov 19, 2019 Infusion Pumps Nationwide View Details →

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Jun 5, 2017 Infusion Pumps Nationwide View Details →

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Jun 5, 2017 Infusion Pumps Nationwide View Details →

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Jun 5, 2017 Infusion Pumps Nationwide View Details →

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Jun 5, 2017 Infusion Pumps Nationwide View Details →

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Jun 5, 2017 Infusion Pumps Nationwide View Details →

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Jun 5, 2017 Infusion Pumps Nationwide View Details →

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sep 28, 2016 Surgical Instruments Nationwide View Details →

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sep 28, 2016 Surgical Instruments Nationwide View Details →

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sep 28, 2016 Surgical Instruments Nationwide View Details →

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sep 28, 2016 Surgical Instruments Nationwide View Details →

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sep 28, 2016 Surgical Instruments Nationwide View Details →

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sep 28, 2016 Surgical Instruments Nationwide View Details →

Greatbatch Medical the manufacturer of the 4F and 5F sheath/dilator components has determined that the products listed in their September 7, 2016 Recall Notification have the potential for the rotating luer to detach from the dilator hub during use. AngioDynamics has confirmed that affected sheath/dilators (Greatbatch Model Numbers 10904-001 and 10904-002) have been included in packaged AngioDynamics Stiffened Micro-Introducer Kits.

Sep 28, 2016 Surgical Instruments Nationwide View Details →

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".

Nov 3, 2015 Surgical Instruments Nationwide View Details →

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".

Nov 3, 2015 Surgical Instruments Nationwide View Details →

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".

Nov 3, 2015 Surgical Instruments Nationwide View Details →

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".

Nov 3, 2015 Surgical Instruments Nationwide View Details →

AngioDynamics Inc., is recalling Micro Introducer Kits and VenaCure EVLT NeverTouch Procedure Kits, because the potential exists that the micro Introducer (sheath/dilator) packaged within the kit could be the incorrect French size.

Apr 30, 2013 Infusion Pumps Nationwide View Details →

AngioDynamics Inc., is recalling Micro Introducer Kits and VenaCure EVLT NeverTouch Procedure Kits, because the potential exists that the micro Introducer (sheath/dilator) packaged within the kit could be the wrong French size.

Apr 30, 2013 Infusion Pumps Nationwide View Details →