The 2D barcodes on these cuvettes contain formatting errors that prevent the analyzing system from correctly reading the expiration date. As a result, the instrument enters an error state and the cuvettes cannot be used.
Becton Dickinson &
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Potential for functionality issue that supports the Over labeling feature and could lead to possible false negative results.
BD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.
One component of the kit was shipped with an expiration date of 30Sep2024 while the Certificate of Analysis (COA) indicated an expiration date of 30Apr2025.
Potential for signal drift in specific lots of BD PCR Cartridges.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
Synapsys software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint. May result in either a false susceptible or false resistant result. A false susceptible result can cause inadequate or incomplete antibiotic treatment due to the use of an ineffective antibiotic, and a false resistant result may limit access to a preferred antibiotic regimen with possible elongation of therapy
Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,
Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,
May intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates.
There is a potential risk that the image colony count does not adequately reflect the actual plate colony count and may lead to inaccurate counts that would impact patient diagnosis and subsequent care.
Assays may produce false negative results due to decreased activity of the enzyme utilized in manufacturing.
Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for fluorescence and if anything is found, the system will abort the panel with the message Fluorescent interference has been detected. Esculin is expected to trigger this fluorescence and test cancellation.
Products were found to have confirmed drug cross contamination prior to release may lead to incorrect, inadequate, or absent antibiotic treatment or delay in results
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Certain vials of the BD BACTEC" Plus Aerobic/F Culture Vials have a labeling issue where duplicate barcode sequence numbers have been identified on more than one vial.
There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.
IFU- Instructions for Use (IFU) package insert and both the IFU packaged with the test kit and the electronic version on the BD website not approved by FDA,specifically describing specimen type, could cause use of the test with invalid specimen types, which may produce false positive or false negative results.Erroneous results can cause a delay in diagnosis or treatment
Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or need for specimen recollection streaking plated media, and inoculating tubes and slides. In Semi-Automated (SA) mode, plates are automatically selected, barcoded, streaked in a pre-configured pattern while the user manually inoculates plates, tubes, and slides. An optional biosafety cabinet on the SA module provides personnel, product, and environmental protection. The InoqulA+" solution is indicated for use in the clinical laboratory.
Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.
Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic
When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.
May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.
BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System
Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment
Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation) and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment
Some N1 and N2 master mix pouches packaged within the lots of BioGX SARS-CoV-2 Reagents for BD MAX System may be mislabeled. This could result in a kit containing two pouches of N1 master mix or two pouches of N2 master mix, instead of one pouch of each. If laboratory personnel do not notice that both pouches contain tubes with the same color foil seal and use two of the same tubes, there is a low potential for a false negative SARS-CoV-2 result. A false negative SARS-CoV-2 result could lead to an increased risk of disease progression for the patient and additional spread of SARS-CoV-2 if the patient is not quarantined.
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD Phoenix Sulfamethoxazole-trimethoprim (SXT) Resistance, false resistant SXT result with S. aureus could lead to a potential loss of access to a preferred antibiotic regimen and potential reduction in effectiveness of treatment of the infection
BD is experiencing elevated rates of potential false positive results from certain customers when using the BD SARS-CoV-2 Reagents for the BD MAX System. False positive test results could lead to a delay in diagnosis and potentially cause the patient to be exposed to COVID-19 if they are isolated with an infected individual.
May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain
May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain
May exhibit increased levels of artifacts which could be difficult to distinguish from gram positive cocci, increased levels of artifacts could result in a false-positive Gram stain
BD Kiestra InoqulA+ Software Version 20.3. IVD automated specimen processing Catalog Number : 447213
An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate
An anomaly is present in InoqulA / InoqulA+ software version 20.3, has the potential to cause a mismatch between a specimen and plate
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.