Canon Medical System, USA

43 recalls Categories: Other Medical Devices, Diagnostic Equipment

It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.

Sep 2, 2025 Diagnostic Equipment View Details →

When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.

Sep 17, 2024 Diagnostic Equipment View Details →

The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

May 10, 2024 Other Medical Devices Nationwide View Details →

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Jan 29, 2024 Other Medical Devices Nationwide View Details →

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Jan 29, 2024 Other Medical Devices Nationwide View Details →

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Jan 29, 2024 Other Medical Devices Nationwide View Details →

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Nov 15, 2023 Diagnostic Equipment View Details →

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Nov 15, 2023 Other Medical Devices View Details →

The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.

Jun 22, 2023 Other Medical Devices Nationwide View Details →

After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.

Sep 29, 2022 Other Medical Devices Nationwide View Details →

Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.

Dec 20, 2021 Other Medical Devices Nationwide View Details →

A software problem has been identified which could result in the diagnostic imaging system not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the diagnostic imaging system failing resulting in rescanning and reinjection of contrast medium.

Nov 11, 2020 Diagnostic Equipment Nationwide View Details →

The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.

May 21, 2021 Diagnostic Equipment Nationwide View Details →

The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.

May 21, 2021 Diagnostic Equipment Nationwide View Details →

The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.

May 21, 2021 Diagnostic Equipment Nationwide View Details →

The MRI system contains a component that does not meet the appropriate flame retardancy-rating which could result in smoke or fire.

May 21, 2021 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.

Feb 5, 2021 Other Medical Devices Nationwide View Details →
Class I - Dangerous

The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.

Feb 5, 2021 Other Medical Devices Nationwide View Details →
Class I - Dangerous

The x-ray irradiation field may shift with respect to image receiving surface displayed on the screen at some C-arm angles.

Feb 5, 2021 Other Medical Devices Nationwide View Details →

A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.

Nov 9, 2020 Diagnostic Equipment Nationwide View Details →

There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.

Oct 8, 2020 Diagnostic Equipment Nationwide View Details →

Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

Mar 3, 2020 Diagnostic Equipment Nationwide View Details →

Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

Mar 3, 2020 Diagnostic Equipment Nationwide View Details →

Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.

Mar 3, 2020 Diagnostic Equipment Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Mar 25, 2019 Other Medical Devices Nationwide View Details →

software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.

Oct 16, 2018 Diagnostic Equipment Nationwide View Details →

When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value. The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.

May 30, 2018 Diagnostic Equipment Nationwide View Details →

It was found when an operator performs a radiography using the wireless flat panel detector (FPD), a message window was displayed on the monitor stating image transmission was not completed and there was no image. It also showed the OK button to reacquire image data form the FPD, and the Cancel button to cancel the reacquisition. When the operator selects the OK button, the same message window appears. The operator then repeated the same operation several times and finally selected the Cancel button to quit the reacquisition mode.

Apr 18, 2018 Other Medical Devices View Details →