The MUSE Administrator Account active directory or local credential could potentially be used from your local network to log into a GE HealthCare MUSE 5 system and potentially manipulate recorded patient data.
GE Medical Systems Information Technologies
Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms.
Use of the blood pressure cuffs could result in inaccurate non-invasive blood pressure measurements.
A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.
CARESCAPE Central Station (CSCS) software version 2.0.2 units may experience unexpected NO COMM (No Communication) and network communication issues after boot-up or system restart.
Device does not produce a visual or audible impedance respiration APN alarm when an impedance respiration apnea event occurs.