Due to an increase in complaint that the percutaneous reference pin would not fit into the patient reference frame or percutaneous pin adapter.
Medtronic Navigation
Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.
Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
Due to a software anomaly which potentially could result in the surgical planning data shifting to an unintended location.
During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury
In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
Due to increasing complaint trend for the Non-Invasive Patient Tracker system where complaints analysis indicates that users are unable to successfully verify their navigation instruments during image-guided surgery.
Defect in sterile pouch seal, incomplete seal, of lighted Retractors, increases the risk of contamination, which can lead to post-operative infection.
Due to a defect in the outer pouch sterile seal
A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.
Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.
There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.
There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]
Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.
Product was incorrectly assembled which could affect navigation accuracy of the device.
Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.
The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.
An incorrect display of the Biopsy Depth Gauge graphical representation in the Guidance View using Cranial software, which resulted in the biopsy of healthy tissue.
Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue.
Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
Potential for Spinal Clamps to be damaged when opened beyond intended limits, either no opening once attached to a patient's spinous process or leaving a component behind in the patient's body following breakage.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
The instruments are not able to pass instrument verification which is performed prior to the use of each instrument.
Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.
Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated Elevate Inserter with the StealthAiR Spine Frame during spine surgical procedures.
Medtronic Navigation, Inc. announces a voluntary field action for the AxIEM Touch N Go Pointer Probe because the patient registration accuracy may compromise navigation accuracy when using the StealthStationΒΏ System with Touch N Go Pointer Probes for patient registration.
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.
Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.
Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.
Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.
Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.
Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.
Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.
Medtronic Navigation, Inc. announces a voluntary field action for the Medtronic Navigation StealthStation Software applications affected by Neurologica BodyTom/CereTom floor-based scanners.
Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user.
Potential for the Visualase Cooled Laser Applicator System (VCLAS) catheters to melt during procedures.
Medtronic Navigation is recalling the Biotex Adapter Kit because it was commercially distributed by BioTex without a cleared premarket [510(k)] from FDA.
Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-armΒΏ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality
Specific use conditions can increase the possibility of a screwdriver tip deforming or breaking during surgery.
Specific use conditions can increase the possibility of a screwdriver tip deforming or breaking during surgery.
The adhesive between the sphere assembly and the biopsy needle shaft may be compromised, causing the sphere assembly to move on the needle shaft. This may result in inaccurate navigation and improper placement of the needles biopsy window in the brain.
The adhesive between the sphere assembly and the biopsy needle shaft may be compromised, causing the sphere assembly to move on the needle shaft. This may result in inaccurate navigation and improper placement of the needles biopsy window in the brain.
Medtronic Navigation is initiating a field correction due to the potential for injury which could occur as a result of cleaning the Medtronic Navigated Cannulated Taps.
Passive Planar Blunt Probe (Lot# 150223) was manufactured with a sharp tip rather than a blunt tip.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
Medtronic Navigation, Inc. has initiated a field correction to their Framelink software version 5.4 which can be used on Medtronic Navigation StealthStation 57 and i7 systems or a Medtronic Planning Station due to the potential of discrepancy between the exam image and the displayed measurement or overlay under certain conditions of use.
Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction
Medtronic is recalling certain Vertex Max Drill Tubes which are used with the StealthStation System due to a defect related to the internal threading for component set up.
A fault in a high voltage relay in the 0-armΒΏ Imaging System might cause the relay to overheat and burn.
Software Version 3.1.1 does not meet navigational accuracy.