Philips Medical Systems Gmbh, DMC

33 recalls Categories: Other Medical Devices, Diagnostic Equipment

Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.

Jul 22, 2020 Other Medical Devices Nationwide View Details →

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Dec 19, 2019 Diagnostic Equipment Nationwide View Details →

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Dec 19, 2019 Diagnostic Equipment Nationwide View Details →