52 recalls β€’ β€’ Categories: Diagnostic Equipment, Other Medical Devices, Infusion Pumps, Implants & Prosthetics, Surgical Instruments

A specific lot of the FlexFlowTM Venous Cannula, 200-200, is being recalled because it is sharper or pointier than expected.

Sep 28, 2017 Diagnostic Equipment Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Infusion Pumps Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Jun 29, 2017 Other Medical Devices Nationwide View Details β†’

Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.

Jun 30, 2017 Other Medical Devices Nationwide View Details β†’

Sorin Group USA, Inc. announces a voluntary field action for the Perceval Sutureless Heart Valve due to updated steps associated with the implantation procedure.

Oct 31, 2016 Implants & Prosthetics Nationwide View Details β†’

On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).

Aug 1, 2014 Other Medical Devices Nationwide View Details β†’

Sorin Group USA, Inc. announces a voluntary field action for the ORCHESTRA Programmer and ORCHESTRA PLUS Programmer because the residual longevity estimate may be less accurate.

Sep 29, 2016 Implants & Prosthetics Nationwide View Details β†’

Sorin Group USA, Inc. announces a voluntary field action for the ORCHESTRA Programmer and ORCHESTRA PLUS Programmer because the residual longevity estimate may be less accurate.

Sep 29, 2016 Implants & Prosthetics Nationwide View Details β†’

Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.

Aug 25, 2016 Diagnostic Equipment Nationwide View Details β†’

Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.

Aug 12, 2016 Diagnostic Equipment View Details β†’

The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.

Aug 12, 2016 Diagnostic Equipment View Details β†’

Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.

Jun 22, 2016 Diagnostic Equipment Nationwide View Details β†’

The CDI System 500 monitor displays the "H/S DISCONNECT AT CUVETTE" error message when the CDI HIS Cuvette does not make a proper connection with the CDI H/S Probe. When this occurs, blood parameter values for HCT, Hgb and S02 do not display.

Jan 19, 2016 Infusion Pumps Nationwide View Details β†’

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Sep 19, 2012 Other Medical Devices Nationwide View Details β†’

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Sep 19, 2012 Infusion Pumps Nationwide View Details β†’

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi

Sep 19, 2012 Infusion Pumps Nationwide View Details β†’

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Sep 19, 2012 Infusion Pumps Nationwide View Details β†’

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determin

Sep 19, 2012 Infusion Pumps Nationwide View Details β†’

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Sep 20, 2012 Infusion Pumps Nationwide View Details β†’

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.

Sep 20, 2012 Infusion Pumps Nationwide View Details β†’

Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.

Sep 14, 2012 Other Medical Devices Nationwide View Details β†’

Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.

Sep 14, 2012 Other Medical Devices Nationwide View Details β†’