Zimmer

422 recalls • • Categories: Implants & Prosthetics, Other Medical Devices, Infusion Pumps, Surgical Instruments, Diagnostic Equipment

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Sep 26, 2025 Implants & Prosthetics Nationwide View Details →

Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.

Sep 26, 2025 Implants & Prosthetics Nationwide View Details →

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

Sep 17, 2024 Infusion Pumps Nationwide View Details →

Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.

Jul 2, 2024 Implants & Prosthetics Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

Mar 20, 2024 Other Medical Devices Nationwide View Details →

Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.

Jun 27, 2023 Other Medical Devices View Details →

Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009

Feb 27, 2023 Implants & Prosthetics View Details →

Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.

Dec 6, 2022 Other Medical Devices Nationwide View Details →

Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.

Dec 6, 2022 Other Medical Devices Nationwide View Details →

Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.

Dec 6, 2022 Other Medical Devices Nationwide View Details →

Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.

Dec 6, 2022 Other Medical Devices Nationwide View Details →

Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.

Dec 6, 2022 Other Medical Devices Nationwide View Details →

Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.

Dec 6, 2022 Other Medical Devices Nationwide View Details →

The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.

Aug 25, 2022 Implants & Prosthetics Nationwide View Details →

It was found through review of open Complaints and stock investigations that a single lot of Zimmer Universal Locking System 3.5 mm Stainless Steel Base and Trays has incorrect silk screening as applied by the supplier. The T-plate catalog numbers noted in the tray are listed as the 00-4946-xxx-xx series (titanium plates); however, the correct numbers should be 00-4936-xxx-xx (stainless steel plat

Aug 11, 2015 Implants & Prosthetics View Details →

Product labeling of these sterile guide wires (external carton label and patient record label) states a 10-year expiration date. Package testing supports a 5-year expiration date for these same guide wires.

Aug 27, 2015 Other Medical Devices Nationwide View Details →

Failed to meet established requirements for sterile barrier integrity due to a breach of the inner cavity. If the outer tray is compromised in addition to the inner tray a periprosthetic infection may occur. The risk associated with a periprosthetic infection is a revision or potential multi-stage revision to treat the infection.

Oct 2, 2014 Implants & Prosthetics Nationwide View Details →

Zimmer is initiating a voluntary recall of 64 lots (752 implants total) of M/L Taper with KinectivĀæ femoral stems and modular necks due higher than allowed cytotoxicity levels found with the product. Reasonable probability of adverse biological response and subsequent revision

May 18, 2015 Implants & Prosthetics Nationwide View Details →

The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.

Apr 29, 2015 Surgical Instruments Nationwide View Details →

Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.

Apr 30, 2015 Implants & Prosthetics Nationwide View Details →

Zimmer, Inc is voluntarily recalling 22 lots of PersonaĀæ Anterior Referencing Sizer with Locking Boom, part number 42-5099-088-10. Devices are being removed from distribution due to nonconforming to print specifications.

Mar 10, 2015 Implants & Prosthetics Nationwide View Details →

Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.

Dec 11, 2014 Implants & Prosthetics Nationwide View Details →

Potential for fracture of the tip of the product. The fractured tip is approximately 4mm x 7mm x 3.5mm in size. Retained tip may result in: pain, soft tissue irritation, soft tissue damage, or increased wear due to contact with the fractured tip; Revision surgery to remove the fractured tip

Nov 25, 2014 Implants & Prosthetics Nationwide View Details →

The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi

Dec 4, 2014 Implants & Prosthetics View Details →

Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.

Aug 26, 2013 Surgical Instruments Nationwide View Details →

Cleaning process validation failure.

Oct 9, 2014 Implants & Prosthetics Nationwide View Details →

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Aug 7, 2014 Implants & Prosthetics Nationwide View Details →

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Aug 7, 2014 Implants & Prosthetics Nationwide View Details →

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Aug 7, 2014 Implants & Prosthetics Nationwide View Details →

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Aug 7, 2014 Implants & Prosthetics Nationwide View Details →

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Aug 7, 2014 Implants & Prosthetics Nationwide View Details →

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Aug 7, 2014 Implants & Prosthetics Nationwide View Details →

Moore & Thompson Hip Stems: FNSTRD HIP 38HDX127ST FNSTRD HIP 40HDX127ST FNSTRD HIP 41HDX127ST FNSTRD HIP 42HDX127ST FNSTRD HIP 43HDX127ST FNSTRD HIP 44HDX127ST FNSTRD HIP 45HDX140ST FNSTRD HIP 46HDX140ST FNSTRD HIP 47HDX140ST FNSTRD HIP 48HDX140ST FNSTRD HIP 49HDX140ST FNSTRD HIP 50HDX152ST FNSTRD HIP 51HDX152ST FNSTRD HIP 52HDX152ST FNSTRD HIP 53HDX152ST FNSTRD HIP 54HDX152ST FNSTRD HIP 55HDX152ST FNSTRD HIP 57HDX152ST FNSTRD HIP 60HDX152ST FNSTRD HIP 63HDX152ST FNSTRD HIP 38HDX140ST FNSTRD HIP 40HDX140ST FNSTRD HIP 41HDX140ST FNSTRD HIP 42HDX140ST FNSTRD HIP 43HDX140ST FNSTRD HIP 44HDX140ST FNSTRD HIP 45HDX152ST FNSTRD HIP 46HDX152ST FNSTRD HIP 47HDX152ST FNSTRD HIP 48HDX152ST FNSTRD HIP 49HDX152ST FNSTRD HIP 50HDX152ST FNSTRD HIP 51HDX152ST HIP PROS 38HDX105ST HIP PROS 40HDX105ST HIP PROS 41HDX105ST HIP PROS 42HDX105ST HIP PROS 43HDX105ST HIP PROS 44HDX105ST HIP PROS 45HDX105ST HIP PROS 46HDX105ST HIP PROS 47HDX105ST HIP PROS 48HDX105ST HIP PROS 49HDX105ST HIP PROS 50HDX105ST HIP PROS 51HDX105ST HIP PROS 52HDX105ST HIP PROS 53HDX105ST HIP PROS 54HDX105ST HIP PROS 55HDX105ST HIP PROS 57HDX105ST HIP PROS 60HDX105ST HIP PROS 63HDX105ST FNSTRD HIP 38HDX165ST FNSTRD HIP 40DX165ST FNSTRD HIP 41HDX165ST FNSTRD HIP 42HDX165ST FNSTRD HIP 43HDX165ST FNSTRD HIP 44HDX165ST FNSTRD HIP 45HDX165ST FNSTRD HIP 46HDX165ST FNSTRD HIP 47HDX165ST FNSTRD HIP 48HDX165ST FNSTRD HIP 49HDX165ST FNSTRD HIP 50HDX191ST FNSTRD HIP 51HDX191ST FNSTRD HIP 52HDX191ST FNSTRD HIP 53HDX191ST FNSTRD HIP 54HDX191ST FNSTRD HIP 55HDX191ST FNSTRD HIP 57HDX191ST FNSTRD HIP 60HDX191ST FNSTRD HIP 63HDX191ST FNSTRD HIP 38HDX165ST FNSTRD HIP 41HDX165ST FNSTRD HIP 43HDX165ST FNSTRD HIP 44HDX165ST FNSTRD HIP 46HDX165ST FNSTRD HIP 48HDX165ST FNSTRD HIP 49HDX165ST FNSTRD HIP 51HDX165ST HIP PROS 38HDX127ST HIP PROS 40HDX127ST HIP PROS 41HDX127ST HIP PROS 42HDX127ST HIP PROS 43HDX127ST HIP PROS 44HDX127ST HIP PROS 45HDX140ST HIP PROS 46HDX140ST HIP PROS 47HDX140ST HIP PROS 48HDX140ST HIP PROS 49HDX140ST HIP PROS 50HDX152ST HIP PROS 51HDX152ST HIP PROS 52HDX152ST HIP PROS 53HDX152ST HIP PROS 54HDX152ST HIP PROS 55HDX152ST HIP PROS 57HDX152ST HIP PROS 60HDX152ST HIP PROS 63HDX152ST HIP PROS 38HDX165ST HIP PROS 40HDX165ST HIP PROS 41HDX165ST HIP PROS 42HDX165ST HIP PROS 43HDX165ST HIP PROS 44HDX165ST HIP PROS 45HDX165ST HIP PROS 46HDX165ST HIP PROS 47HDX165ST HIP PROS 48HDX165ST HIP PROS 49HDX165ST HIP PROS 50HDX191ST HIP PROS 51HDX191ST HIP PROS 52HDX191ST HIP PROS 53HDX191ST HIP PROS 54HDX191ST HIP PROS 55HDX191ST HIP PROS 57HDX191ST HIP PROS 60HDX191ST HIP PROS 63HDX191ST

Class I - Dangerous

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Jul 22, 2014 Implants & Prosthetics View Details →

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Jul 22, 2014 Implants & Prosthetics View Details →

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Jul 22, 2014 Implants & Prosthetics View Details →

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Jul 22, 2014 Other Medical Devices View Details →

Trauma Tubes and Plates: TUBE & SCP PLT 90DX150LG TUBE & SCP PLT 90DX205LG TUBE & SCP PLT 90DX255LG TUBE & SCP PLT 95DX150LG TUBE & SCP PLT 95DX205LG TUBE & SCP PLT 95DX255LG TUBE&SCP PLT 130DX8H TUBE&SCP PLT 130DX10H TUBE&SCP PLT 130DX12H TUBE&SCP PLT 130DX14H TUBE&SCP PLT 135DX8H TUBE&SCP PLT 135DX10H TUBE&SCP PLT 135DX12H TUBE&SCP PLT 135DX14H TUBE&SCP PLT 140DX8H TUBE&SCP PLT 140DX10H TUBE&SCP PLT 140DX12H TUBE&SCP PLT 140DX14H TUBE&SCP PLT 145DX8H TUBE&SCP PLT 145DX10H TUBE&SCP PLT 145DX12H TUBE&SCP PLT 145DX14H TUBE&SCP PLT 150DX6H TUBE&SCP PLT 150DX8H TUBE&SCP PLT 150DX10H TUBE&SCP PLT 150DX12H TUBE&SCP PLT 150DX14H VERSA-FX II SC TUBE PLT 90DX6H VERSA-FX II SC TUBE PLT 90DX8H VERSA-FXII SC TUBE PLT 90DX10H VERSA-FXII SC TUBE PLT 90DX12H VERSA-FXII SC TUBE PLT 90DX14H VERSA-FXII SC TUBE PLT 95DX6H VERSA-FXII SC TUBE PLT 95DX8H VERSA-FXII SC TUBE PLT 95DX10H VERSA-FXII SC TUBE PLT 95DX12H VERSA-FXII SC TUBE PLT 95DX14H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE VERSA-FX II STD TUBE 130DX8H VERSA-FX II STD TUBE 130DX10H VERSA-FX II STD TUBE 130DX12H VERSA-FX II STD TUBE 130DX14H VERSA-FX II STD TUBE 135DX8H VERSA-FX II STD TUBE 135DX10H VERSA-FX II STD TUBE 135DX12H VERSA-FX II STD TUBE 135DX14H VERSA-FX II STD TUBE 140DX8H VERSA-FX II STD TUBE 140DX10H VERSA-FX II STD TUBE 140DX12H VERSA-FX II STD TUBE 140DX14H VERSA-FX II STD TUBE 145DX8H VERSA-FX II STD TUBE 145DX10H VERSA-FX II STD TUBE 145DX12H VERSA-FX II STD TUBE 145DX14H VERSA-FX II STD TUBE 150DX8H VERSA-FX II STD TUBE 150DX10H VERSA-FX II STD TUBE 150DX12H VERSA-FX II STD TUBE 150DX14H VERSA-FX II KEYLESS TB 130DX8H VERSA-FX II KEYLESS TB 135DX8H VERSA-FX II KEYLESS TB 140DX8H VERSA-FX II KEYLESS TB 145DX8H VERSA-FX II KEYLESS TB 150DX8H VERSA-FX KEYLESS 90D X 8HOLE VERSA-FX KEYLESS 90D X 11HOLE VERSA-FX KEYLESS 95D X 8HOLE VERSA-FX KEYLESS 95D X 11HOLE VERSA-FX KEYLESS 95D X 14HOLE

Class I - Dangerous

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Jul 22, 2014 Other Medical Devices View Details →

Trauma Large Screws: CANN BONE SCREW 7.0MMX100MMLG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0MMX105MMLG CANN BONE SCREW 7.0X110LG CANN BONE SCREW 7.0X110LG CANN BONE SCREW 7.0MMX110MMLG CANN BONE SCREW 7.0X115LG CANN BONE SCREW 7.0X115LG CANN BONE SCREW 7.0MMX115MMLG CANN BONE SCREW 7.0X120LG CANN BONE SCREW 7.0X120LG CANN BONE SCREW 7.0MMX120MMLG CANN BONE SCREW 7.0X125LG CANN BONE SCREW 7.0MMX125MMLG CANN BONE SCREW 7.0X130LG CANN BONE SCREW 7.0MMX130MMLG RECON SCR LGE 5.5MMDX105MM RECON SCR LGE 5.5MMDX110MM RECON SCR LGE 5.5MMDX115MM RECON SCR LGE 5.5MMDX120MM RECON SCR LGE 5.5MMDX125MM RECON SCR LGE 5.5MMDX130MM ITST ANTI-ROT SCR 6.5MMDX105MM ITST ANTI-ROT SCR 6.5MMDX110MM PERI. SCR 3.5MM X105MML PERI SCR. 6.5MM X 100MM L PERI SCR. 6.5MM X 140MM L PERI SCR. 6.5MM X 100MM L PERI SCR. 6.5MM X 100MM L 3.5MM CORT SCR X 105MM SELFTAP 3.5MM CORT SCR X 110MM SELFTAP 5.5MM X 100MM CANN LKNG SCREW Z NAIL 6.0 X 105 CANC SCREW PT Z NAIL 6.0 X 110 CANC SCREW PT Z NAIL 6.0 X 115 CANC SCREW PT Z NAIL 6.0 X 120 CANC SCREW PT Z NAIL 6.0 X 125 CANC SCREW PT Z NAIL 6.0 X 130 CANC SCREW PT Z NAIL 6.0 X 135 CANC SCREW PT Z NAIL 6.0 X 140 CANC SCREW PT Z NAIL 6.0 X 105 CANC SCREW FA Z NAIL 6.0 X 110 CANC SCREW FA Z NAIL 6.0 X 115 CANC SCREW FA Z NAIL 6.0 X 120 CANC SCREW FA Z NAIL 6.0 X 125 CANC SCREW FA Z NAIL 6.0 X 130 CANC SCREW FA Z NAIL 6.0 X 135 CANC SCREW FA Z NAIL 6.0 X 140 CANC SCREW FA Z NAIL 6.0 X 105 CANC SCREW PT Z NAIL 6.0 X 110 CANC SCREW PT Z NAIL 6.0 X 115 CANC SCREW PT Z NAIL 6.0 X 120 CANC SCREW PT Z NAIL 6.0 X 125 CANC SCREW PT Z NAIL 6.0 X 130 CANC SCREW PT Z NAIL 6.0 X 135 CANC SCREW PT Z NAIL 6.0 X 140 CANC SCREW PT Z NAIL 6.0 X 105 CANC SCREW FA Z NAIL 6.0 X 110 CANC SCREW FA Z NAIL 6.0 X 115CANC SCREW FA Z NAIL 6.0 X 120CANC SCREW FA Z NAIL 6.0 X 125 CANC SCREW FA Z NAIL 6.0 X 130 CANC SCREW FA Z NAIL 6.0 X 135 CANC SCREW FA Z NAIL 6.0 X 140 CANC SCREW FA 3.5X105 CORT SCR 3.5 X 105 CORT SCR SELF TAP 3.5X110 CORT SCR 3.5 X 110 CORT SCR SELF TAP 3.5X115 CORT SCR 3.5 X 115 CORT SCR SELF TAP 3.5X120 CORT SCR 3.5 X 120 CORT SCR SELF TAP 3.5X125 CORT SCR 3.5X130 CORT SCR 3.5X135 CORT SCR 3.5X140 CORT SCR 3.5X145 CORT SCR 3.5X150 CORT SCR 4.5 X 105 CORT SCR SELF TAP 4.5 X 110 CORT SCR SELF TAP 4.5 X 120 CORT SCR SELF TAP 6.5X105 CANC SCR, FULLY THDED 6.5X105 CANC SCR, 16MM THD 6.5X105 CANC SCR, 32MM THD 6.5X110 CANC SCR, FULLY THDED 6.5X110 CANC SCR, 16MM THD 6.5X110 CANC SCR, 32MM THD 6.5X115 CANC SCR, 32MM THD 6.5X120 CANC SCR, 32MM THD

Class I - Dangerous

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Jul 22, 2014 Other Medical Devices View Details →

Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to final manufacturing under Part 00-5926-014-01; lot 62455332. A forged lap is a defect that forms when metal folds over itself during forging.

Jul 30, 2014 Implants & Prosthetics Nationwide View Details →

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Jul 22, 2014 Other Medical Devices View Details →

Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.

Jun 10, 2013 Implants & Prosthetics Nationwide View Details →

Persona (TM) The Personalized Knee System PS Tibial Articular Surface Provisional Top Nonsterile TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 6-9 CD TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 3-5 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 6-9 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 10-11 EF TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 6-9 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 GH TOP TASP PS L 10-12 J TOP TASP PS L 10-12 J TOP TASP PS L 10-12 J TOP TASP PS L 10-12 J TOP TASP PS L 10-12 J TOP TASP PS L 10-12 J TOP TASP PS L 10-12 J TOP

Class I - Dangerous

Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.

Jun 10, 2013 Implants & Prosthetics Nationwide View Details →