There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Diagnostic Equipment
π₯ Medical Devices β’ 6,627 recalls
Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
COBAS INTEGRAΒΏ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
COBAS INTEGRAΒΏ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
Edwards Lifesciences has received a limited number of customer reports regarding an inter-lumen leak between the cardioplegia lumen and pressure lumen in the device.
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
CELL-DYN Emerald analyzer, REF 09H39-01. The CELL-DYN Emerald is an automated hematology analyzer intended for in-vitro diagnostic use in the clinical laboratory.
Abbott Laboratories
There is a potential for the device to generate Quality Control (QC) low or out-of-range low for parameters RBC and PLT.
McKesson Consult Diagnostic Immunochemical Fecal Occult Blood Test, Part numbers 178-CK30T, 178-CT30, 178ΒΏTBSO, 178-CM25, 178-CK50, 178-CM50, 178-CT50, 178ΒΏCK30
Cypress Medical Products
Products were denied Clinical Laboratory Improvement Amendments (CLIA) waiver approval by the FDA, due to the packaging being too similar to domestically sourced products under the same branded label.
The firm issued an updated instructions for use to reduce the potential for phrenic nerve injury.
The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.