The invasive pressure visual and audible alarms may not activate.
Diagnostic Equipment
π₯ Medical Devices β’ 6,627 recalls
ACS-NT, Model 78108 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body,or extremities.
Philips Electronics North America
Potential risk for helium gas inside the MR examination room during a magnet quench
BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
VITROS Immunodiagnostic Products TSH Reagent Pack
Clinical Diagnostic Systems
VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.
IntelliVue X3 Patient Monitor.
Philips Electronics North America
The NBP measurement of Intellivue X3 Patient Monitor shows intermittently only mean values instead of the diastolic and systolic blood pressure values. This is caused by a falsely detected NBP cuff, which is leading to wrong internal NBP setting. Furthermore, occasionally the Monitor shows the Check Touch Input message and the monitor is inoperable with the touch interface.
There is possibly a difference between the measured dioptric value of the subjective measurement mode and the objective measurement mode on ARK-1s.
The manufacturer received complaints indicating that the device's central handle holder detaches easily from the handle holder ring on the light head during clinical use. In the event of such detachment, particles from the handle holder ring could fall into the sterile field during a procedure.
BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
Product has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower torso along with the arm housing
BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.
The device may process the same sample with two different sample IDs and sample information while using the single tube loader, which has the potential for erroneous results due to the mis-identification.
The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use.
if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).
if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).
Potential for loose or missing CO2 diffusers which could result in a non-functioning mask.
The Oncomine Dx Target Test may report erroneous results when used off-label for samples other than non-small cell lung cancer (NSCLC) tumor specimens.
A database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity PACS-IW. There is a potential that one or more image series (i.e. all images within an image set) may be missing from an exam without indication to the user. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR Thorax exam with 1 image per series.
The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube. Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.
Increased amount of residual blood present on the top of stopper well after venipuncture