a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.
Diagnostic Equipment
🏥 Medical Devices • 6,627 recalls
Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical chemistry systems.
Randox Laboratories
According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice
During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure + first primary barrier are closed the signal will be transmitted to the generator for activating X-ray. During intended use only with this footswitch X-ray should be activated.
The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.
Incorrect expiration date was listed on the label. Correct expiration date was November 30, 2016. Product was labeled with expiration date of April 30, 2018.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.
Siemens Healthcare Diagnostics
Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8¿C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.
Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.