An issue with the cancel functionality used during atypical infusion programming to cancel user inputted values. An infusion may start that is greater than or less than the hospital established limits for the specific medication.

May 13, 2015 Diagnostic Equipment Nationwide View Details β†’

Potential for false positive Chlamydia trachomatis (CT) results and Neisseria gonorrhoeae (GC) results when processing the CT Qx and GC Qx Assays on the BD Viper LT Instrument.

Nov 7, 2014 Diagnostic Equipment View Details β†’
Class I - Dangerous

The A-plane compensator is made of Teflon which will become fragile due to X-ray exposure passing through it. The thinnest part of the compensator is subject to radiation degradation and as a result has been found to fail prematurely. Failure is indicated by pin or ring artifact in the image

Mar 8, 2014 Diagnostic Equipment View Details β†’

Integra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Catheters because they may have been incorrectly assembled. Specifically, the o-ring in the bolt adapter may be in the wrong location and as a result, there is a potential CSF/blood could leak from the bolt adapter.

May 27, 2015 Diagnostic Equipment Nationwide View Details β†’

complaints of system malfunction and unintended, sudden movement at start up. No injuries reported.

Apr 30, 2015 Diagnostic Equipment Nationwide View Details β†’

Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.

Apr 6, 2015 Diagnostic Equipment Nationwide View Details β†’

The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

Mar 6, 2015 Diagnostic Equipment Nationwide View Details β†’

The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

Mar 6, 2015 Diagnostic Equipment Nationwide View Details β†’

Software anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no pretreatment when the sample was programmed.

Apr 6, 2015 Diagnostic Equipment Nationwide View Details β†’