Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
Diagnostic Equipment
🏥 Medical Devices • 6,627 recalls
Liquid Waste Bottle Kit, Catalog#10386287. The Immulite 2000 and 2000 XPi systems are continuous random-access instruments that perform chemiluminescent immunoassays. These instruments use serum, plasma, or urine samples for in vitro diagnostic testing.
Siemens Medical Solutions Diagnostics
The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
High invalid rate and an increased risk of false negative results with certain lots of the Alere i Influenza A & B kit.
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
Liquid Waste Bottle, Catalog#10291587. The Immulite 2000 systems are continuous random-access instruments that perform chemiluminescent immunoassays. These instruments use serum, plasma, or urine samples for in vitro diagnostic testing.
Siemens Medical Solutions Diagnostics
The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.
Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Medical Systems. The device is a diagnostic imaging system for fixed installations that combines Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI).
Philips Medical Systems (Cleveland)
It was noticed that the spectroscopy voxel map did not align with the anatomical images of the phantom. This resulted in incorrect orientation of the Chemical Shift Imaging (CSI) data.
Customers complained about under-recovery of non-Roche controls and discrepant low patient results with certain Homocysteine reagent lots. This negative bias could, in the worst case, lead to inaccurately low Homocysteine results. An elevated level of Homocysteine is considered an important risk factor in the assessment of peripheral vascular disease. Falsely low values could lead to a delay of di
GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were expelled. There were no incidents or injuries reported as a res
The probability and histogram data within the Melafind's device displayed user interface lacks approval (not included in the PMA supplement).
Richard Allan Scientific (RAS) is initiating a recall of Mucolytic agent and Mucolexx following a customer complaint reporting that a lab had found fungal contamination in 2 lots of Mucolexx.
Integra LifeSciences Corporation is recalling Integra's Camino ICP Monitor (CAM 02) & Licox PtO2 Monitor ( LCX02 & LCX02R) due to complaints that the audible speaker malfunctions.
Integra LifeSciences Corporation is recalling Integra's Camino ICP Monitor (CAM02) & Licox PtO2 Monitor ( LCX02 & LCX02R) due to complaints that the audible speaker malfunctions.
Mucolytic Agent 1 pint/473 ml; 1 gallon/3.8 l Product Usage: A mucoliquefying preservative designed for use in the preparation of mucoid cytology specimens. Mucolexx & Mucolytic Agent liquefies mucus and allows for optimal cytologic diagnostic results.
Richard-Allan Scientific Company
Richard Allan Scientific (RAS) is initiating a recall of Mucolytic agent and Mucolexx following a customer complaint reporting that a lab had found fungal contamination in 2 lots of Mucolexx.
When EPIQ 7 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).
When EPIQ 5 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).
Integra LifeSciences Corporation is recalling Integra's Camino ICP Monitor (CAM 02) & Licox PtO2 Monitor ( LCX02 & LCX02R) due to complaints that the audible speaker malfunctions.
The test kits have Out-of-Range-Low (OORL) Flucytosine QC results for Candida krusei ATCC 6258
The test kits have Out-of-Range-Low (OORL) Flucytosine QC results for Candida krusei ATCC 6258.