smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.
Smith & Nephew
Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.
A single lot of GENESIS II FEMORAL FIVE DEGREE VALGUS BUSHING due to a laser mark labeling error. The left valgus bushing was incorrectly laser marked as right and vice versa.
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1003
Kelyniam Global
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.
There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1004
Kelyniam Global
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.
There is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.
Reports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and death.
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
Kelyniam Custom Skull Implant (CSI), Model no. PEEK-IM1002
Kelyniam Global
Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
ATTUNE Revision CRS Femoral RT SZ 5 Cemented, Catalog No. 150440205, Femoral Revision Implant, Knee
DePuy Orthopaedics
There may be burrs on the extraction hole threads.
Gentle Threads, Interference Screw; Model No. 905604, for soft tissue reattachment procedures in the ankle/foot and knee.
Zimmer Biomet
Specific lots of the device were overexposed during EtO sterilization. This may lead to revision due to infection from loss of sterile barrier during over-exposure cycle.
RESTORIS MCK TIBIAL BASEPLATE-RM/LL used in unicompartmental knee arthroplasty and knee replacements.
Mako Surgical
Incorrect product and/or label. Sizing is labeled incorrectly.
ATTUNE Revision CRS Femoral LT SZ 5 Cemented, Catalog No. 150440105, Femoral Revision Implant, Knee
DePuy Orthopaedics
There may be burrs on the extraction hole threads.
Bard Medical Division is recalling the BARD¿ BrachySource¿ Iodine125 Radioactive Seeds in Mick¿ Applicators and BARD¿ BrachySource¿ Iodine125 Radioactive Seeds Loose in Vial with Sterile Spacers. Because the incorrect dosage level radioactive seeds were mixed up and sent to two hospitals. One Hospital ordered BrachySource¿ Iodine125 0.36 mCi radioactive seeds but instead received BrachySource¿ Iodine125 0.28 mCi radioactive seeds and vice versa for the other hospital affected.
POWERPORT ClearVUE Slim Implantable Port with Smooth Septum and Attachable 8F Polyurethane Open-Ended Single-Lumen Venous Catheter Custom Kit, REF CP00005
Bard Peripheral Vascular
The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.
The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.
POWERPORT ClearVUE isp Implantable Port With Smooth Septum and Attachable 8F Polyurethane Open-Ended Single-Lumen Venous Catheter Custom Kit, REF CP00004
Bard Peripheral Vascular
The Patient ID Card packaged with the POWERPORT ClearVUE Implantable Port with Smooth Septum contains incorrect port flushing instructions.