Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
New User Manual with additional instructions for use and a hose and mask adapter provided by the device manufacturer. This voluntary recall is intended to help reduce potential health risks of exposure to ozone gas after the use of a SoClean device.
Cordis became aware that the label for the carton box containing 10 units is incorrect for some boxes of two lots of MynxGrip. The carton box from lot F2322902 (Item MX5021, MynxGrip 5F) as labeled as lot F2322903 (MX6721, MynxGrip 6/7F) and product from lot F2322903 (MX6721, MynxGrip 6/7F) was labeled as F2322902 (Item MX5021, MynxGrip 5F). The primary label on individual devices is correct.
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
ROi CPS manufactured and distributed medical convenience kits with an irrigation solution component which was subsequently recalled by the component manufacturer.
One lot of Part Number 373025, XstarΒΏ Kojo Safety Slit Knife 2.5mm 45ΒΏ bevel up, incorrectly contains a bevel down blade instead of a bevel up blade.
Firm is recalling all kits containing recalled lots of Nurse Assist Sterile 0.9% Normal Saline. Nurse Assist identified product in distribution for which the required sterility assurance level (SAL) of 10^-6 cannot be guaranteed. Risks to patients if product is used include blood stream infections, urinary tract infections, or respiratory infections; in the worst-case scenario, there is risk of sepsis or death.
Keyprint Keysplint Hard Clear 1 kg resin bottles were inadvertently packaged with the wrong Instructions for Use. Keysplint Hard Clear were placed inside a zip lock bag along with KeySplint Soft Instructions for Use (IFU).