Intera 1.5T Power/Pulsar
Philips North America
Potential component failure in the Gradient Coil could product smoke and/or fire.
🏥 Medical Devices • 15,809 recalls
Philips North America
Potential component failure in the Gradient Coil could product smoke and/or fire.
Philips North America
Potential component failure in the Gradient Coil could product smoke and/or fire.
Sentinel CH SpA
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
Instrumentation Laboratory
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
Medtronic Neuromodulation
Units distributed in Korea and Turkey that did not have the correct firmware installed to support the user interface in the local language.
Certain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not being made with Di(2-ethylhexyl) phthalate (DEHP); however, Avanos has determined that the tubing components of the product did contain DEHP.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Atrium Medical
Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
Philips North America
This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).
Physio-Control
Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
Physio-Control
Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
Specific Regard kits contain the Gentige Chest Drain Products being recalled by Gentige. The Instructions for Use (IFU) for the chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains.
A complaint investigation showed that the bulk pellets (lot 794-142) were contaminated with Aspergillus flavus.
Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
Physio-Control
Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
MEDLINE INDUSTRIES, LP - Northfield
This recall is being issued due to items within the kit have expired prior to the overall kit expiration date listed on the insert.
MEDLINE INDUSTRIES, LP - Northfield
This recall is being issued due to items within the kit have expired prior to the overall kit expiration date listed on the insert.
Physio-Control
Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
MEDLINE INDUSTRIES, LP - Northfield
This recall is being issued due to items within the kit have expired prior to the overall kit expiration date listed on the insert.