Other Medical Devices

🏥 Medical Devices 15,809 recalls

Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.

Sep 12, 2023 Other Medical Devices View Details →
Class I - Dangerous

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Oct 12, 2023 Other Medical Devices Nationwide View Details →

Allura Xper IGTS Fixed Systems

Philips North America

Class I - Dangerous

Philips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.

Oct 4, 2023 Other Medical Devices View Details →
Class I - Dangerous

Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

Sep 15, 2023 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Philips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.

Oct 4, 2023 Other Medical Devices View Details →