One lot of Tournier Perform Reversed Lateralized Baseplates were found to have been manufactured with 6mm spacers instead of 3mm spacers. This nonconformance prevents the central screw from being able to thread into the baseplate.
Other Medical Devices
🏥 Medical Devices • 15,809 recalls
Specific lots of the Dual Lumen HydroPICC Kits were provided with the incorrect revision of the Instructions For Use (IFU) which do not include the following warnings statement: HydroPICC Dual Lumen Catheter is for adult use only. There is a potential risk to health should the HydroPICC Dual Lumen Catheter be used in an inadequately sized vessel. There have been no reported adverse events to date.
Specific lots of the Dual Lumen HydroPICC Kits were provided with the incorrect revision of the Instructions For Use (IFU) which do not include the following warnings statement: HydroPICC Dual Lumen Catheter is for adult use only. There is a potential risk to health should the HydroPICC Dual Lumen Catheter be used in an inadequately sized vessel. There have been no reported adverse events to date.
Functional MR V1.0
OLEA MEDICAL
When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause file contents and associated series descriptions to not match, which could lead to a misinterpretation of clinical data.
The thumb strap of the wrist and thumb orthosis may contain latex that is not declared in the product labeling.
Procedure pack: Presource PBDS Cat. PLMHCSBHF, Hermann Hospital MHHS, Kit, C Section B MHHS, Sterile.
Cardinal Health 200
Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).
Multichem P, Part Code 08P90-10
TECHNO-PATH MANUFACTURING
Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.
MAZOR X robotic guidance system, Model TPL0059, with software versions 4.2.2 or 5.0.1
Mazor Robotics
Software update
Procedure pack: Presource PBDS, Cat. PEMHEEESR, Southeast Hospital MHHS, Kit, ENT, SE, Sterile.
Cardinal Health 200
Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).
2008T HD SYS. W/O CDX BLUESTAR
Fresenius Medical Care Holdings
Potential PCBA leaching from tubing of hemodialysis machines.
REVERSE PROSTHESIS SHOULDER SYSTEM, SCREW, LOCKING BONE, RSP, 5mmX30mm long, REF: 506-03-130
Encore Medical
Reverse shoulder prosthesis package that should contain 5mmx30 screws could contain incorrect 3.5mmx30 screws. Use of incorrect screws could result in surgical delay and inadequate fixation.
Constellation Vision System, REF: 8065751150
Alcon Research
Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.
2008T Hemodialysis SYS, with CDX
Fresenius Medical Care Holdings
Potential PCBA leaching from tubing of hemodialysis machines.
CellQuicken Analyzer (Smart-Watch and Software), RoyalVibe Ultrasound Generator, Brainvibe, Envirovibe Water Restructuring Pad, RoyalVibe Therapy Balls
RoyalVibe Health
Ultrasound devices were not authorized, cleared, or approved by FDA, and thus their performance characteristics have not been adequately established. There is a possibility of customers' reliance on marketing claims which could cause the user to forgo medical treatment resulting in health consequences.
Procedure packs: (1) Presource Pain Tray, Catalog SBAHFPTAS2, Advanced Surgery Center SAR, Sterile. (2) Presource Block Tray, Catalog SNE37ANMOF, Medical Center Odessa.
Cardinal Health 200
Packs contain recalled components (equipment drapes, table cover, and surgical room turnover (SRT) kits).
Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12
TECHNO-PATH MANUFACTURING
Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.
2008T HD SYS. W/O CDX W/BIBAG BLUESTAR
Fresenius Medical Care Holdings
Potential PCBA leaching from tubing of hemodialysis machines.
MED-810A Zemits NDPrime Laser products
Advance-Esthetic
Advance-Esthetic LLC failed to comply with the applicable standards regarding Electronic Product Radiation Control (EPRC) regulations associated with the laser product performance standards and certification. This Recall provides correction by the firm regarding Zemits NdPrime Yag Tattoo Removal Laser. The product requires update or is missing the following in part: remote interlock connector, safety interlocks, manual reset mechanism, beam attenuator, tests for determination of compliance, certification, class IV designation and warning, labels, and labeling.
3T Heater Cooler System, REF: 16-02-85
LivaNova Deutschland GmbH
The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device.
Access Thyroglobulin Calibrators
Beckman Coulter
It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.