2D Perfusion, release R1.0.x, R1.1.x, R1.2, and R1.2.1
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
🏥 Medical Devices • 15,203 recalls
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.
There is a potential for breakage of the small screw used to affix the anti-backout plate.
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Beckman Coulter
The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This introduces a risk of the positive bias not being detected by QC with the potential that clinical correlation with other thyroid function tests will be inconsistent requiring additional testing; however the clinical assessment and treatment decisions are not made in isolation on a single test result especially, in the setting of thyroid function testing. False positive test results could lead to a euthyroid patient being interpreted clinically as having elevated thyroid function (hyperthyroid) or a hypothyroid patient being within the normal range. Such misdiagnoses may lead to inappropriate treatment of a euthyroid patient with anti-thyroid medications or hypothyroid patient not getting sufficient treatment since their free T3 test result demonstrates them to be in the euthyroid range.
Avanos Medical
Incorrect expiration date.
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
AGILENT TECHNOLOGIES INC./US
Distributed sample collection kit with an unapproved instruction for use.
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Philips North America
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Philips North America
If the system experiences an overvoltage / power surge it can cause a short circuit condition within the Main- cabinet. If this issue occurs, the user may experience an electrical fire within the cabinet, see or smell smoke emitting from the cabinet, or hear a popping sound coming from the cabinet.
Philips North America
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users
Philips North America
If the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear; as a result there is a potential that the Ceiling Suspension of the DigitalDiagnost C50 system may descend inadvertently if power to the system is interrupted, and may result in injuries to patients and users
Coloplast Manufacturing US
A decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.