Automated dispensing cabinet drawer firmware may have inadvertently been downgraded to an earlier version of drawer firmware during new device installations or device replacement upgrades, which may lead to an increased frequency of drawer failures, which could increase the likelihood of patients experiencing delays in receiving medications.
Other Medical Devices
🏥 Medical Devices • 15,203 recalls
TEMA Axle Small, REF 1590.15.010 (additional information is pending receipt of the firm's product label/labeling)
Limacorporate S.p.A
Due to the incorrect component, that is not cleared to be marketed in the U.S., being included in its elbow prosthesis.
TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.213S
Synthes (USA) Products
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Model: 650705550001
Stryker Medical Division of Stryker
Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.
Deep Inspiration Breath Hold (DIBH) functionality may not work as specified when based on a standard-workflow Treatment Template
Amicus Exchange Kit Therapeutics. Component of Amicus Separator multiprocedural apheresis platform
Fenwal
Potential for centrifuge packs to develop a stress leak for certain lots of Amicus MNC Apheresis kit and Amicus Exchange kit (Therapeutic Kits) on the Amicus Separator.
TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 170MM - STERILE. Intended for temporary fixation and stabilization of proximal femur fractures. Part Number: 04.037.212S
Synthes (USA) Products
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Recreo Hair Growth Laser Comb
OMM Imports, Inc.
The firm distributed the laser products without affixing the warning, aperture, certification, and identification labels. In addition, the firm failed to provide safety information, reproductions and locations of the labels, and a warning statement in the user manual.
An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a remote probability of delayed results or reporting erroneous patient results. The subject nut bands were found to not meet the firm's specifications for hardness and had the incorrect material grain direction. Affected instruments were manufactured from December 28, 2017 to October 19, 2022 or had their nut bands replaced during the same period.
Fusion Bioline Vascular Graft, Part number M00201503046B0
Maquet Cardiovascular
One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.
Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.
Arial mobile application, model #54630 and #54640 for Android devices downloaded from the Google Play Store as part of the Arial Emergency and Nurse Call system.
Securitas Healthcare
The mobile application for Android devices downloaded from the Google Play Store do not load per specifications and function on the Android devices.
Potential for product manifold to crack during use on patients, possibly causing leaks in the respiratory circuit and can lead to the patient receiving inadequate ventilation.
TFN-Advanced Femoral Nail 12MM/130 DEG TI CANN TFNA 170MM STERILE. Intended for temporary fixation and stabilization of proximal femur fractures Part Number: 04.037.242S
Synthes (USA) Products
TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length
Product is labeled with an incorrect expiration date.
Heartsine Samaritan PAD: SAM 350P semi-automatic defibrillator, SAM 360P fully automatic defibrillator, SAM 500P semi-automatic defibrillator with CPR Advisor
HeartSine Technologies
Automated external defibrillators were potentially distributed with the incorrect language configuration, which can lead to a delay in therapy or no therapy.
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.
Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.
Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.
Linet Spol. S.r.o.
Labeling being updated to assure user is securing the siderail to a locked position to prevent potential patient fall and injury.