Other Medical Devices

🏥 Medical Devices 15,809 recalls

Class I - Dangerous

This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.

Mar 15, 2022 Other Medical Devices Nationwide View Details →

The Virus Counter is an automated laboratory instrument that enables rapid quantification of intact viruses in solution. This instrument is not a medical device. The virus quantification is used in process development and manufacturing of various biotechnology products such as vaccines, gene therapy and personalized immunotherapy. Device is a Class 1 laser flow cytometer.

Mar 11, 2022 Other Medical Devices Nationwide View Details →

Artis zee floor MN, Model Number 10094142

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q ceiling, Model Number 10848281

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q zeego, Model Number 10848283

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q.zen floor, Model Number 10848353

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee biplane, Model Number 10094141

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →
Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee biplane MN, Model Number 10094143

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan

Mar 7, 2022 Other Medical Devices Nationwide View Details →

Artis Q floor, Model Number 10848280

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee ceiling, Model Number 10094137

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q.zen biplane, Model Number 10848355

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q biplane, Model Number 10848282

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q.zen ceiling, Model Number 10848354

Siemens Medical Solutions USA

Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →