Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Other Medical Devices
π₯ Medical Devices β’ 15,809 recalls
A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.
The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.
Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031
Philips North America
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.
The device has similarities in design to another device that the firm recently issued a field correction on due to instances of tip detachment and embolization.
If a new primary image set is selected while the cine loop is running, the primary image set will be displayed as both primary and secondary image set in all side-by-side views. This will also be true for any new patient or case opened while the cine loop is running
CIRRUS HD-OCT, Models 5000 AngioPlex and 6000 AngioPlex
Carl Zeiss Meditec
Optic nerve head angiography scan to be turned off due to its distribution without pre-market clearance.
The boxes used by the shipper were not validated for use during transport.
Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.
Acorn 130 Stairlift, Brooks 130 Stairlift, Acorn 180 Stairlift, and Acorn Outdoor Stairlift
Acorn Stairlifts
The design of the seats did not consistently have the adequate strength required to support the customers weight with continued use.
Product was labelled with the incorrect manufacturing and distribution dates.
MEDLINE Micro-Kill Bleach Germicidal Bleach Wipes
MEDLINE INDUSTRIES, LP - Northfield
Micro-Kill Bleach Wipes contain out of specification (low) levels of Sodium Hypochlorite.
Alinity m Integrated Reaction Units (IRU)
Abbott Molecular
There is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.
Digital Angiography System Bransist safire Generator Model D150GC-40
Shimadzu Medical Systems
Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
Collagen Plugs, CP3, 0.3mm, CP4, 0.4mm, CP5, 0.5 mm, Sterile, EO, Rx Only, CE Dry eye treatment
Lacrimedics
The sterile pouch seal may contain channels that could affect the seal integrity and potentially compromise sterility.
BUTLER Clear Dip, 12 bottles/case
Sunstar Americas
Product lacks premarket clearance.
The labeling for the Microbiology Brush identifies the device as a Bronchial Microbiology Brush. The device is only cleared for gastrointestinal use and is not cleared for bronchial use.
Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.
Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.