Other Medical Devices

πŸ₯ Medical Devices β€’ 15,809 recalls

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

Feb 26, 2021 Other Medical Devices View Details β†’

The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.

Jan 29, 2021 Other Medical Devices Nationwide View Details β†’

Operating Table System

Getinge Group Logistics America

Class I - Dangerous

There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.

Feb 11, 2021 Other Medical Devices Nationwide View Details β†’

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

Feb 26, 2021 Other Medical Devices View Details β†’
Class I - Dangerous

Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.

Feb 25, 2021 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.

Feb 25, 2021 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.

Feb 25, 2021 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.

Feb 25, 2021 Other Medical Devices Nationwide View Details β†’

JUDKINS PACK

Cardinal Health 200

Class I - Dangerous

Component Part Number: 19537LBL (LBL, SHEET OF 39, ASSORTED, STDSZ, MULTI W/BLK INK, WATERPROOF) is being reported as intermittently mixed with an incorrect medication label sheet throughout the specific PresourceΒΏ kit SANHDJTVCF Judkins Pack lot # 525067.

Feb 11, 2021 Other Medical Devices View Details β†’

Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.

Feb 24, 2021 Other Medical Devices Nationwide View Details β†’

SOMATOM systems-issues in with software syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a, may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Delay in diagnosis or patient rescan may occur. Sporadic software errors during interventional workflows may also result

Mar 3, 2021 Other Medical Devices Nationwide View Details β†’

Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory

Feb 11, 2021 Other Medical Devices View Details β†’

Maquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.

Jan 29, 2021 Other Medical Devices Nationwide View Details β†’