i-STAT cTnI cartridge
Abbott Point Of Care
i-STATΒΏ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.
π₯ Medical Devices β’ 15,809 recalls
Abbott Point Of Care
i-STATΒΏ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.
Villa Sistemi Medicali S.P.A.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)
Invia Foam Dressing Kit with FitPad could have holes in the sterile packaging.
STANLEY BLACK & DECKER
The products are labeled as Class II lasers. However, the level of laser radiation emitted by the distributed products may exceed the limit (1 mW) established for Class II lasers. These products would be more appropriately classified into laser Class IIIa (IEC Laser Class IIIR).
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
There is a potential performance issue as detected by an under-recovery of sample results during routine testing.
When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
Reports of pressure relief manifold venting gas below the standard pressure.
The firm has become aware that the Snapshot Trackers were not manufactured in accordance with specifications and could potentially lead to navigational inaccuracies, which may result in patient injury, e.g. nerve damage or neurological deficit/dysfunction
There is a potential performance issue as detected by an under-recovery of sample results during routine testing.
Medtronic Xomed
Routine sterilization dose does not meet the required Sterility Assurance Level.
Diagnostica Stago
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.
When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information Center (CIC) systems were identified to have vulnerabilities to a cyber-attack.
Reports of pressure relief manifold venting gas below the standard pressure.
Reports of pressure relief manifold venting gas below the standard pressure.
Medtronic Xomed
Routine sterilization dose does not meet the required Sterility Assurance Level.
Diagnostica Stago
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.