Other Medical Devices

🏥 Medical Devices 15,809 recalls

BrightView X , Model Number 882478

Philips Medical Systems (Cleveland)

Class I - Dangerous

An issue with the handcontroller may result in either spontaneous uncommanded motions or continued motion after the buttons were released.

Jun 10, 2019 Other Medical Devices Nationwide View Details →

AXIOM Artis BA, Model Number 5904656

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

RX Daytona, Model Nos. RX4040 and RX4041

Randox Laboratories, Limited

Class I - Dangerous

There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.

Jun 24, 2019 Other Medical Devices View Details →

AXIOM Artis BC, Model Number 5904649

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

RX Imola, Model Nos. RX4900

Randox Laboratories, Limited

Class I - Dangerous

There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.

Jun 24, 2019 Other Medical Devices View Details →

AXIOM Artis FC, Model Number 5904433

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

AXIOM Artis dBC, Model Number 5917054

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

AXIOM Artis MP, Model Number 5904466

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

AXIOM Artis TC, Model Number 7728350

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.

Jul 18, 2019 Other Medical Devices Nationwide View Details →

AXIOM Artis dFC, Model Number 7412807

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

Artis zee biplane MN, Model Number 10094143

Siemens Medical Solutions USA

Class I - Dangerous

A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.

Jun 12, 2019 Other Medical Devices Nationwide View Details →

Compatibility issue between the software and Microsoft updates to Windows 7, 8, and 10. The compatibility issue causes the software to be unusable which could result in a delay of reporting results.

Sep 17, 2018 Other Medical Devices View Details →