There is a possibility that the seal of the sterile pouch may be compromised, thereby increasing the risk of infection.
Other Medical Devices
π₯ Medical Devices β’ 15,809 recalls
Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002
Stryker GmbH
When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.
Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764
Terumo Cardiovascular Systems
Screws on a TerumoΒΏ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system
Visualase Thermal Therapy System, Software: Visualase Software versions 3.1.1 - 3.2, Product #: 9735542
Medtronic Navigation
Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue.
Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763
Terumo Cardiovascular Systems
Screws on a TerumoΒΏ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.
The products failed the filed specification of 1.0kPa for Burst Pressure.
The Gentle Thread PLGA Round Head, 7x20mm and the Gentle Thread PLGA Full Thread 10x30 have been commingled.
4.5 Fr x 40 cm single-lumen PICC Kit, Product Code CDC-44041-VPS2, 5.5 Fr double-lumen PICC Kit, Product Code CDC-44052-VPS2
Arrow International
The products may contain the incorrect catheter. Kits that should contain a 4.5 Fr single-lumen catheter may contain a 5.5 Fr double-lumen catheter. Kits that should contain a 5.5 Fr double-lumen catheter may contain a 4.5 Fr single-lumen catheter. This issue may cause a delay in treatment while a replacement catheter is obtained.
Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems
Accuray Incorporated
the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System
AIRO Mobile CT System Model # MobiCT-32
Mobius Imaging
The Tube Current Modulation feature (Modulated Scans) is not working in AIRO systems with software version 2.0.0.0, and operators would not be able to detect this fault until after a scan is completed.
ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000
Siemens Medical Solutions USA
Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).
2008T HEMODIALYSIS SYSTEM W/BIBAG, Part Number 190766, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.
Fresenius Medical Care Renal Therapies Group
In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.
2008T HEMODIALYSIS SYS, WITH CDX, Part Number 190713, Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138 Product Usage: Indicated for acute and chronic dialysis therapy.
Fresenius Medical Care Renal Therapies Group
In the hemodialysis machine during the cleaning/disinfection program, when there is a power failure or interruption, the dialysis program button should be greyed out when the machine is powered up and a mandatory rinse should be required. However, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine enters the dialysis program even though the dialysis program button is disabled and the mandatory rinse has not been performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.
The temperature sensor catheters may be defective in that they will show a lower body temperature measurement than the actual temperature of the patient.
The product was mislabeled with the incorrect label.
The Gentle Thread PLGA Round Head, 7x20mm and the Gentle Thread PLGA Full Thread 10x30 have been commingled.
DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries
Pro-Med Instruments Gmbh
The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.
sterEOS workstation
EOS Imaging
3D projections may be incorrect when the pair of images used is a secondary capture generated from a series containing more than two images, where one of the images has been manually flipped. Projection errors may lead to errors in calculation of 3D clinical parameters when adjusting the position, size and shape of the 3D objects used. Incorrect clinical parameters can be a contributing factor in incorrect clinical decisions, or sub-optimal surgical interventions.
Freedom EV0-2 with any version of EVOware and EVOware Sample Tracking Add-On V1.0 or higher
Tecan US
There is a software issue that can potentially impact sample identification and traceability under specific conditions when using EVOware Add On Software: Sample Tracking -Multi Instrument V1.0 or higher, Part Number 30019971.
Artis Q Interventional Fluoroscopic X-Ray System, Model Numbers 10848280, 10848281, 10848282, 10848283, 10848354 & 10848355
Siemens Medical Solutions USA
Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).