Other Medical Devices

πŸ₯ Medical Devices β€’ 15,809 recalls

Class I - Dangerous

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

May 24, 2018 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

Possible degradation of an internal electronic component over time which may cause the XLTEK EMU 40EX Breakout box to fail. It may feel hot and plastic case may start melting.

May 15, 2018 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

May 24, 2018 Other Medical Devices Nationwide View Details β†’

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

May 2, 2018 Other Medical Devices View Details β†’

Optima CT 520

GE Healthcare

Class I - Dangerous

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

May 24, 2018 Other Medical Devices Nationwide View Details β†’

Optima CT 540

GE Healthcare

Class I - Dangerous

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

May 24, 2018 Other Medical Devices Nationwide View Details β†’

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

May 2, 2018 Other Medical Devices View Details β†’

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

May 2, 2018 Other Medical Devices View Details β†’
Class I - Dangerous

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

May 24, 2018 Other Medical Devices Nationwide View Details β†’

ELITech Clinical Systems SAS initiated this recall because some users of ENVOY 500 ISE CALIBRATOR Kit (Part number: 55117), for Envoy 500 systems are observing trouble while calibrating with some vials. Users contacted the firm when they received the messages "ISE OUT OF REPRODUCIBILITY" or "ISE SLOPE OUT OF RANGE" (i.e., failed calibration). This has resulted in delay in obtaining patient results until the calibrator lot is replaced.

Apr 18, 2017 Other Medical Devices View Details β†’