🏨

Other Medical Devices

πŸ₯ Medical Devices β€’ 15,809 recalls

Product labeled as GVL 2 Stat (blades), lot GC32825, was physically packaged with the incorrect size device (GVL 1 Stat (blades)).

Mar 3, 2017 Other Medical Devices Nationwide View Details β†’

Artis zee, Angiographic x-ray system

Siemens Medical Solutions USA

Class I - Dangerous

Software error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system functions properly until the next reactivation. Following reactivation, the system no longer initializes as intended.

Mar 28, 2017 Other Medical Devices View Details β†’

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Mar 22, 2017 Other Medical Devices Nationwide View Details β†’

The On/Off functionality in the LCSU 4 is controlled by electronic components, one of which might be damaged.

Jan 8, 2017 Other Medical Devices Nationwide View Details β†’

Potential for poly tip of the mallet to come off during use. Potential for foreign material in the surgical field or delay in surgery. Reminder for customers that the poly tip must be removed for cleaning and sterilization to assure proper cleaning and sterilization of the entire mallet.

Feb 20, 2017 Other Medical Devices Nationwide View Details β†’

Alcon is initiating a Voluntary Medical Device Removal because the product supplier has stated that there is potential for sterile packaging to be compromised.

Mar 27, 2017 Other Medical Devices Nationwide View Details β†’

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Mar 22, 2017 Other Medical Devices Nationwide View Details β†’
Class I - Dangerous

During an antivirus program scan of the Eye Station or Eye Care PACS, the antivirus program may detect a newly captured image to be an unwanted file and delete the image permanently.

Apr 4, 2016 Other Medical Devices Nationwide View Details β†’