Other Medical Devices

🏥 Medical Devices 15,809 recalls

CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the sample result. Orthos diluents used to dilute samples containing CRP are manufactured using human plasma based fluids that contain endogenous CRP. Without accounting for the CRP in the diluent and removing its contribution from the result, the reported result will include the CRP present in both the sample and diluent, possibly leading to positively biased results.

Oct 26, 2016 Other Medical Devices Nationwide View Details →

There is a problem with the mating feature between the plate and screw which could cause the plate opening not to interfere enough with the plate and head of the screw.

Nov 22, 2016 Other Medical Devices Nationwide View Details →

The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.

Sep 16, 2016 Other Medical Devices Nationwide View Details →

BD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer¿ Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and rear barrels upon activation of the safety feature, which retracts the needle.

Aug 8, 2016 Other Medical Devices Nationwide View Details →

neoBLUE blanket LED Phototherapy System distributed since release of the product in 2011 is recalled due to customer complaints regarding discoloration/ degradation of fiberoptic bundle at connection of pad and box.

Oct 28, 2016 Other Medical Devices Nationwide View Details →
Class I - Dangerous

GE Healthcare has become aware of a potential safety issue where certain Avance CS2, Avance and Amingo anesthesia devices can transition to a System Malfunction state if the lower storage drawer containing the optional large tray insert accessory closed with an abnormally high amount of force.

Nov 3, 2016 Other Medical Devices Nationwide View Details →

GF Health Products, Inc., in conjunction with OKIN America, has initiated a voluntary field correction affecting the OKIN America "DUOMAT 3" Head/Foot motor used in the manufacture of our Lumex Patriot Full Electric Bed Model US0458/US0468 and Patriot LX Full Electric Bed Model US6000. The failure mode is directly related to contaminated material in the plastic housing used to produce the DUOMAT 3 motor. Contaminated motor housings may crack resulting in a motor that will no longer actuate the head section per the intended design. If the head section were to fall it is unlikely but possible an injury could occur. No injuries have been reported.

Sep 30, 2016 Other Medical Devices Nationwide View Details →

Firm became aware that professional samples of Ceramax are not meeting stability standards set for the product.

Nov 21, 2016 Other Medical Devices Nationwide View Details →

The reason for the recall is that the inner carton box of the Rolled Probe Cover PC0907NB has the incorrect expiration date (2015-10-01) on the label while the individual pouch labels and outer shipper have the correct expiration date (2018-10-01). The manufacturing date was inadvertently used as the expiration date on the inner carton box labels.

Nov 30, 2016 Other Medical Devices View Details →

Accuray has become aware of a potential safety issue involving unexpected treatment robot motion when removing an accessory. This issue is limited to certain CyberKnife¿ Systems with software version 1 0.6.

Nov 22, 2016 Other Medical Devices View Details →