Ultra PC% Cabinet Mount Package G, Model #36600, Serial # 13593, 13594, 13595, 13596, 13597, 13598, 13599, 13600, 13602, 13605, 13610, 13611, 13612, 13615, 13616, 13617, 13621, 13622, 13624, 13625, 13627, 13628, 13633, 13634, 13635, 13636, 13637, 13638, 13639, 13640, 13642, 13643, 13644, 13647, 13650, 13669, 13675, 13673, 13678, 13679, 13680, 13683, 13685, 13687, 13691, 13693, 13694, 13695, 13696, 13697, 13701, 13702, 13705, 13706, 13709, 13710, 13711, 13724, 13725, 13728, 13729, 13735, 13739, 13743, 13746, 13750, 13751, 13753, 13754, 13755, 13756, 13760, 13762, 13766, 13774, 13775, 13777, 13781, 13782, 13783, 13785, 13786, 13788, 13792, 13794, 13796, 13799, 13800, 13803, 13804, 13809, 13810, 28298, 28299, 28304, 28305, 28306, 28307, Ultra PC% Cabinet Mount Pkg I, Model #36800, Serial # 13601, 13603, 13607, 13608, 13609, 13613, 13618, 13620, 13623, 13626, 13629, 13630, 13631, 13632, 13641, 13645, 13646, 13648, 13649, 13651, 13652, 13653, 13654, 13655, 13656, 13657, 13658, 13659, 13660, 13661, 13665, 13666, 13667, 13668, 13671, 13672, 13674, 13676, 13677, 13684, 13688, 13689, 13698, 13700, 13707, 13723, 13727, 13730, 13732, 13734, 13741, 13742, 13757, 13758, 13759, 13761, 13762, 13763, 13764, 13765, 13767, 13769, 13789, 13790, 13791, 13795, 13797, 13801, 13802, 13806, 13808, 28300, 28302, 28303, Ultra PC% Cabinet Mount Pkg G, Model#36600-NS, Serial# 13805, 13807, 13811, 13812. Ultra PC% Cabinet Mount Pkg H, Model#36800-NS, Serial# 13744, 13745,13752, 13773, 13774, 13778, 13779, 13780 Product Usage: Conscious Sedation Unit Analgesia Gas Machine.

Accutron

Class I - Dangerous

Accutron is recalling the Flowmeter (Analgesia Gas Machine) because when it flows Nitrous Gas without Oxygen gas being turned on, it functions in an incorrect manner.

Apr 29, 2012 Other Medical Devices Nationwide View Details β†’

Zimmer is initiating a recall of the Cemented Tibial Drill for the Zimmer Persona Personalized Knee System due to three reports being received as of August 29. 2012 for tibial cortex perforation during preparation of the tibial bone. The investigation into the reports that have been received found that the perforations are occurring while using the Cemented Tibial Drill. The perforations have o

Sep 7, 2012 Other Medical Devices Nationwide View Details β†’

Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw

Aug 23, 2012 Other Medical Devices Nationwide View Details β†’

The recall was initiated because Beckman Coulter has confirmed that the Coulter PrepPlus may experience intermittant reagent level sense failures where the instrument may not produce a warning that a reagent vial is low or out of reagent.

Jan 20, 2012 Other Medical Devices Nationwide View Details β†’

The values reported for PCO2 by BP7561 B-Lac cassettes are biased at high concentrations when used at altitudes above 2800 feet (853 meters).

Jul 3, 2012 Other Medical Devices Nationwide View Details β†’

Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat

Sep 20, 2012 Other Medical Devices Nationwide View Details β†’

Terumo Cardiovascular Systems (CVS) received two reports of the false back flow alarms for the TerumoΒΏ Advanced Perfusion System 1 when there was no tubing in the flow sensor. Terumo CVS' investigation found that the voltage level on a circuit board was improperly set in a population of flow modules for Terumo System 1, the sensor operates correctly if the tubing is installed and it is highly unli

Sep 10, 2012 Other Medical Devices Nationwide View Details β†’

Product was distributed with an incorrect expiration date. Product was labeled with an expiration date of 2013-04 and should have been labeled with 2012-09.

Aug 13, 2012 Other Medical Devices Nationwide View Details β†’

Johnson & Johnson Healthcare Products (JJHP) is voluntarily recalling this product to address FDA concerns that the product should have been cleared under a separate medical device application. This voluntary recall is not being conducted on the basis of any consumer safety or product performance issues related to K-Y LIQUIBEADS. JJHP received a Warning Letter on May 23, 2012 indicating that thi

Jul 2, 2012 Other Medical Devices Nationwide View Details β†’

KCI has issued a medical device correction for the BariAir Therapy System Hand Pendent. The BariAir Therapy System is not designed to allow the patient to place the bed into the patient exit position. KCI has received complaints indicating that patients used the hand pendent to inadvertently place the BariAir Therapy System into a pseudo-exit position. In a small number of instances, this allow

Sep 17, 2012 Other Medical Devices Nationwide View Details β†’

Varian has identified an anomaly with the Eclipse Treatment Planning where an outdated assigned Hounsfield Unit [HU] value can be used for dose calculation when the clock on the Eclipse Client Workstation is not synchronized with the Database/System Server. Treatment of the patient using these values can therefore lead to under or over-dose.

Aug 17, 2012 Other Medical Devices Nationwide View Details β†’

Asahi KASEI, REXEED(TM)- 13 LX, HEMODIALYZER dialysis filter

Asahi Kasei Medical Co., Ltd., Okatomi Plant

Class I - Dangerous

Asahi became aware of a negative trend for rate of the adverse events that may be associated with a manufacturing change to certain lot numbers of the products, though there is no sufficient information to date to conclusively link these adverse events to the products.

Sep 21, 2012 Other Medical Devices View Details β†’