Update to IFU provides a manual deployment workaround method to help mitigate potential harms related to partial stent deployment.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
a. Rapid Test Kit Coag-Sense Professional Blood Coagulation Test Prothrombin Time Test (PT/INR) Whole Blood Sample 50 Tests-IVD. Part No. 0356-50. b. Coagulation Control Kit Coag-Sense PT / INR Test (Prothrombin Time Test) Low Level / High Level 10 Strips Part No. 03P69-10.
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
FISHER SCIENTIFIC Fisherbrand SWAB, ROUND BOTTOM CLR (50/PK)FSHSCI Catalog # 1490712
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
a. ABX PENTRA CREATININE 120CP JAFFE 120/130. Ref: A11A01933. b. Reagent General Chemistry Creatinine Model Number: 1220001933
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
EVIS EXERA III Bronchovideoscope
Aizu Olympus Co.
The recalled unit was utilized in a veterinary endoscopy procedure in advance of being assigned to healthcare facilities as a loaner.
Baxter DISCPAC Syringe Tip Caps, Self-Righting, Luer Slip, Dark Blue: a) 25 pack REF H938693025, b) 100 pack REF H938673100
Baxter Healthcare
There is a potential of the packaging not maintaining a sterile barrier for the tip caps.
MedSource IV Start kit, 8inch Extension Set IV START KIT, W/SITE LOC/8" EX100/CS Model Number: MS-80047
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Device does not meet regulatory requirements of stability while stationary, which requires mechanical equipment, other than fixed mechanical equipment, that is intended to be used on the floor or on a table shall be permanently marked with a clearly legible warning of this risk
Perifix LOR Syringe Perifix 8 mL Pouch Luer Lock Tip Without Safety Model Number: 332152
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
There is a potential for leaking valves on ports 1 to 4 in certain lots of valve sets.
Nobel Biocare N1 TiUltra TCC NP 3.5x13mm
Nobel Biocare
Due to the presence of a burr which may cause an aspiration risk to patient during surgical procedure.
Excelsior Heparin Sodium, Porcine, Preservative Free 100 U / mL Solution Prefilled Syringe, Partial Fill 5 mL Fill in 10 mL Syringe Model Number: 513604
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
CENTURION MEDICAL PRODUCTS, LP Centurion CATH KIT, FML W/LUBE 8FR (50/CS) CNTURN Catalog # CKF105L
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
BECKMAN COULTER, INC. a. COULTER DxH Diff Pack REAGENT, DIFF PK FP DXHCOULTR, Catalog #628020. b. COULTER DxH Cleaner REAGENT, CLEANER FP DXH 10LCOULTR Catalog #628023. c. DxH 500 Series Lyse LYSE, VERSA SOL DXH 500 SERIES(0.5L/KT) Catalog # B36846. d. DxH 500 Series Cleaner CLEANER, COULTER CLENZ DXH 500SERIES (0.5L/KT) Catalog # B36868
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
MAGELLAN DIAGNOSTICS LeadCare II Blood Lead Test Kit Rapid Test Kit LeadCare II General Chemistry Blood Lead Test Whole Blood Sample 48 Tests Model Number: 70-6762
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Baxter DISCPAC Syringe Tip Caps, Self-Righting, Luer Slip, Green: a) 25 pack, REF H93869025, b) 100 pack, REF H938674100, c) 100 pack, REF H93869100
Baxter Healthcare
There is a potential of the packaging not maintaining a sterile barrier for the tip caps.
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
Stryker SWANSON Flexspan Finger Joint Implant W/O Grommets Silicone, Size: 00, REF 4700020, Sterile.
Wright Medical Technology
An incorrect sizing label was applied to the carton packaging.
a. Medefil Normal Saline IV Flush Syringe NORMAL SALINE FLUSH, SYR 10ML/12ML (60/BX 16BX/CS), Model Number: MIS-1130. b. MEDEFIL INC McKesson SALINE IV FLUSH, SYR NORMAL 10ML/12ML .9% (60/BX 16BX/CS) Model Number: 191-MIS-1130.
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
a. MERIT MEDICAL SYSTEMS Aspira Dressing Kit Aspira Model Number: 4991503. b. MERIT MEDICAL SYSTEMS Aspira Pleural / Peritoneal Drainage Kit Aspira 1000 mL Sterile Model Number: 4992301.
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.