Dreamtome RX Cannulating Sphincterotome-DREAMTOME 44-20MM/260CM Material Number: M00584040
Boston Scientific
Sterility of device is compromised due to a sterile barrier breach
Medical equipment, implants, and diagnostic devices
Boston Scientific
Sterility of device is compromised due to a sterile barrier breach
Illegal importation of COVID-19 AG Test kits into the U.S. The COVID-19 Ag Test are not approved, cleared or authorized for distribution within the U.S.
Boston Scientific
Sterility of device is compromised due to a sterile barrier breach
Wright Medical Technology
An incorrect screw is packaged in the product, which contains a locking screw instead of a non-locking screw.
Boston Scientific
Sterility of device is compromised due to a sterile barrier breach
Boston Scientific
Sterility of device is compromised due to a sterile barrier breach
The device has similarities in design to another device that the firm recently issued a field correction on due to instances of tip detachment and embolization.
This lot may have an elevated risk of leakage from the Air-Inlet Filter of the device. Leakage of the Air-Inlet Filter may increase the risk of decreasing the dose of medication delivered to the patient, cause delays in treatment due to the need to replace the device or may cause contamination of the fluid path that can lead to bloodstream infection.
Ortho-Clinical Diagnostics
The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.
Semi-automatic defibrillators shipped out by Stryker without battery and electrode packs.
individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result
Bolder Surgical
A potential manufacturing/assembly defect could result in the generator not meeting IEC 60601 standards for medical electrical equipment, and it can potentially, under specific conditions, lead to the delivery of unintended energy to the patient.
The PCA 4000 may not fully cure EnvisionTEC dental resins to desired product specifications.
Kits were labeled with incorrect expiration dates.
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.
A risk for product contamination was identified for the affected product lot. If used, there is an increased potential for SARS-CoV-2 false positive results when using the product.
Lusys Laboratories
COVID-19 test kits (Antibody Rapid Test kit and Antigen Rapid Test Kit) are not authorized, cleared, or approved for marketing and/or distribution in the U.S.
The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.
Philips North America
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.