Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.
There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.
There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.
Stryker received two (2) complaints for units from specific lots of Inner and Outer Dilators not fitting together properly during a surgical procedure. Some dilators from these lots and from other lots were subsequently determined to be affected by a manufacturing nonconformance.
There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.
Microbiologics QC Sets and Panels. BD MAX" CT/GC/TV Verification and Control Panel. Containing High, Medium and Low Level Controls.
Microbiologics
Real-time shelf life testing failed at 24 months
The sterile pouch seal may contain channels that could affect the seal integrity and potentially compromise sterility.
Brasseler USA, K-WIRE, Double Trocar Smooth, REF KM172-26-35S
Brasseler USA, Medical L.L.C.
One lot of product was distributed in unsealed packaging
There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.
NavitrackER Kit A: Knee, Model Number 20-8000-000-07,
Orthosoft, Inc.
The product was released for distribution without passing sterilization results, which could potentially lead to insufficient sterility of the product. The highest severity risk is infection leading to surgical intervention.
LYFO-DISK, Quality control kit for culture media, sold as: LYFO-DISK Enterococcus faecalis derived from ATCC 51299
Microbiologics
This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
MEDPOR Titan OFW - MTB - Right - 1.0mm, Catalog number 81032
Stryker Leibinger GmbH & Co. KG
Inside the packaging of one reported item a non-conforming product was found. Contrary to the design, the barrier layer was manufactured on the inferior side of the implant versus the superior side.
Helix Elite" Molecular Standards: Inactivated Standard. Inactivated Influenza A/B and Respiratory Syncytial Virus
Microbiologics
Real-time shelf life testing failed at 24 months
Digital Angiography System Shimadzu Trinias Generator Model: D150GC-40
Shimadzu Medical Systems
Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.
There is a potential to leak from the catheter where it connects to the drain bag
There is a potential to leak from the catheter where it connects to the drain bag
May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts
IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software
Natus Neurology
Tile kits that allow one to install and situate a video camera within a patient environment are not fitting properly in some domestic institution ceiling tile grids that have a thinner style grid frame. The ceiling tile may suddenly fall from the ceiling and cause a safety risk to the patient or end user.
Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of inadvertently selecting the incorrect coil size for the procedure potentially causing some harms like vessel damage, an inability to treat with said device and potential rupture of the aneurysm.