KWIK-STIK, Candida parapsilosis derived from ATCC 22019, REF 0726P. Packaged 2 units per pack. IVD controls.
Microbiologics
Potential that products packaged for lot 726-57-10 were packaged in pouches labeled as 323-108-7.
Medical equipment, implants, and diagnostic devices
Microbiologics
Potential that products packaged for lot 726-57-10 were packaged in pouches labeled as 323-108-7.
The firm has received six (6) complaint reports from customers indicating difficulty removing the balloon protector sleeve from the catheter (e.g., excessive force required to remove the protector sleeve). If the balloon is damaged while removing the protector sleeve and the catheter is subsequently used in a patient procedure, patient harm may potentially occur, including perforation of vessels, foreign body in patient, embolism, risks associated with surgical intervention, or delay to treatment/therapy.
MEDLINE INDUSTRIES, LP - Northfield
Product stored incorrectly in temperature controlled setting instead of refrigeration.
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
Siemens Medical Solutions USA
Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.
The European version of the DiaTrust COVID-19 Ag Rapid Test is not approved for marketing in the U.S. but was illegally distributed in the U.S. The packaging and labeling of the European test kit is different from the US test kit and is NOT authorized for distribution in the US.
Empowered Diagnostics
COVID test kits were offered for sale and distributed to consumers in the United States without marketing approval, clearance, or authorization from FDA.
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.
The Dry Heat Indicator Labels packaging may contain TTS Indicator Labels instead of Dry Heat Indicator Labels.
Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
Smith & Nephew
The affected screwdrivers are out of specification and may not mate with the Surfix Alpha Lock screw
Empowered Diagnostics
COVID test kits were offered for sale and distributed to consumers in the United States without marketing approval, clearance, or authorization from FDA.
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
MEDLINE INDUSTRIES, LP - Northfield
Product stored incorrectly in temperature controlled setting instead of refrigeration.
The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.
Baxter Healthcare
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.