Ondal AC2000 (PDI) Television Mounting System
Class II - ModerateWhat Should You Do?
- Check if you have this product: (215 distributed to 5 hospitals)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
Recall Details
- Company:
- Unknown Manufacturer
- Reason for Recall:
- The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
- Classification:
- Class II - Moderate
Products that might cause a temporary health problem, or pose a slight threat of a serious nature.
- Status:
- ongoing
- Hazard:
- The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.
- Remedy:
- Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.
Product Images
Product Information
Full Description:
The mounting systems were sold for use in hospital Intensive Care Units. The units are off-white and have multiple joints between the ceiling mount and the TV mount at the end of the arm system.
Product Codes/Lot Numbers:
(215 distributed to 5 hospitals)
Official Source
Always verify recall information with the official CPSC source:
View on CPSC.govCPSC Recall Number: 12742
Related Recalls
Bamamon LED Projecting Finger Light Toys
Unknown Manufacturer
None reported.
Costway Convertible High Chairs
Unknown Manufacturer
The convertible high chairs violate the mandatory safety standard for high chairs because they do not contain a passive restraint system and a child can slip through the opening between the feeding tray and seat and become entrapped or fall, posing a risk of serious injury or death.
CIYTI Light-Up Princess Costumes
Unknown Manufacturer
The costumes violate the mandatory safety standard for toys because the button cell batteries can be accessed easily by children, posing a deadly ingestion hazard. If button cell or coin batteries are swallowed, the ingested batteries can cause serious injuries, including internal chemical burns, and death.