Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI/DI 10837461002611, Lot Numbers: 432-24 and higher
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Beckman Coulter, Inc.
Reason for Recall:
Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.

Product Codes/Lot Numbers:

UDI/DI 10837461002611, Lot Numbers: 432-24 and higher

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0093-2025

Related Recalls

It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.