CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI/DI 20653160222979, Lot Code 2022050690
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
MEDLINE INDUSTRIES, LP - Northfield
Reason for Recall:
One lot of chest tube insertion trays, containing a silk suture component that is restricted to 1 cycle of sterilization, was processed though 2 cycles.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
completed

Product Information

Full Description:

CENTURION SnagFree Chest Tube insertion Tray, Reorder CHT545

Product Codes/Lot Numbers:

UDI/DI 20653160222979, Lot Code 2022050690

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0106-2023

Related Recalls

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.

Jul 25, 2025 Other Medical Devices View Details →

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Flexicare BritePro Solo Fiber Optic Laryngoscope Handles due to the failure of the light to illuminate as intended.

Jul 25, 2025 Other Medical Devices View Details →