Monaco - Radiation Treatment Planning used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software Version 5.10.01
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Elekta, Inc.
- Reason for Recall:
- Unintended update of Dose and MU and Incorrect Assignment of Bolus.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Monaco - Radiation Treatment Planning used to make treatment plans for patients with prescriptions for external beam radiation therapy.
Product Codes/Lot Numbers:
Software Version 5.10.01
Distribution:
Distributed in: CA, IN, MI, MO, TX, WI
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0181-2016
Related Recalls
Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound
Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iViewGT / XVI detector arms. When the detector arm is fully deployed there is no risk of uncontrolled extension.