3.0mm Headless Compression Screw, Long Thread 18mm, Part Number 04.226.118 DePuy Synthes 3.0mm Headless Compression Screws are intended for fixation of intra-articular and extra-articular fractures and nonunions of small bones and small bone fragments.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Numbers: L734474
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Synthes (USA) Products LLC
- Reason for Recall:
- The product inside the package does not match the label. The product was labeled as long thread screws that have a 6mm thread length but were actually short thread screws with a 4mm thread length.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
3.0mm Headless Compression Screw, Long Thread 18mm, Part Number 04.226.118 DePuy Synthes 3.0mm Headless Compression Screws are intended for fixation of intra-articular and extra-articular fractures and nonunions of small bones and small bone fragments.
Product Codes/Lot Numbers:
Lot Numbers: L734474
Distribution:
Distributed in: US, AR, AZ, FL, IL, NM, OH, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0238-2019
Related Recalls
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 46MM / XL25/ STER. Product Code: 04.045.346TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Synthes (USA) Products
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Synthes (USA) Products
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.
STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010
Synthes (USA) Products
Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay