Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- NeuroLogica Corporation
- Reason for Recall:
- When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.
Product Codes/Lot Numbers:
Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F
Distribution:
Distributed in: NC, OH, MA, MO
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0263-2016
Related Recalls
The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.
GM60A. Digital Diagnostic Mobile X-Ray System.
NeuroLogica
Potential for the anti-fall system of the device arm to fail.
Mobile x-ray systems moving arm frame welding issue which presents the potential of bodily harm due to fall of the arm.