Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 WMTS Smart-hopping (1.4GHz) Product Number: 865350 Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 USA only SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    SW Revisions: B.05.28, B.05.29, and B.05.32
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Electronics North America Corporation
Reason for Recall:
Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 WMTS Smart-hopping (1.4GHz) Product Number: 865350 Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 USA only SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals

Product Codes/Lot Numbers:

SW Revisions: B.05.28, B.05.29, and B.05.32

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0291-2017

Related Recalls

There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.

Nov 15, 2018 Diagnostic Equipment View Details →

A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.

Oct 10, 2018 Diagnostic Equipment Nationwide View Details →

Automated external defibrillators may not fully meet IPx5 water ingress specification. The device may fail to function should water intrusion occur.

Oct 10, 2018 Implants & Prosthetics Nationwide View Details →