lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: (01)00813502011371(10)1430951(17)210331(21)192148A; Serial Number: 192148A
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abiomed, Inc.
- Reason for Recall:
- This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI
Product Codes/Lot Numbers:
UDI: (01)00813502011371(10)1430951(17)210331(21)192148A; Serial Number: 192148A
Distribution:
Distributed in: GA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0345-2020
Related Recalls
Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.