smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442024, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot number 15GM08498C
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Smith & Nephew, Inc.
- Reason for Recall:
- Devices, incorrectly labeled as a VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, contained a VLP MINI-MOD 2.0MM STOUT PLATE 8 HOLE and vice versa
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
smith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442024, QTY: (1), STERILE R Bone plates and bone screws from the Smith & Nephew Variable-Angle Locking Mini-Fragment Plating System (VLP MINI-MOD) are intended for use in internal fixation of small bones and small bone fragments and non-load bearing stabilization and reduction of bone fragments in long bones.
Product Codes/Lot Numbers:
Lot number 15GM08498C
Distribution:
Distributed in: US, MI, GA, OH
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0396-2018
Related Recalls
META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690
Smith & Nephew
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.
The Bipolar assembly contains an oversized Retainer Ring.