CombiDiagnost R90; Model Number: 709031;

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: 709031; UDI-DI: 00884838101456; Serial Number: 100001385, 100001402, 100001409, 100001412, 100001415, 100001416, 100001417, 100001424, 100001426, 100001432, 100001449, 100001460, 100001464, 100001476, 100001480, 100001483, 100001488;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Medical Systems DMC GmbH
Reason for Recall:
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

CombiDiagnost R90; Model Number: 709031;

Product Codes/Lot Numbers:

Model Number: 709031; UDI-DI: 00884838101456; Serial Number: 100001385, 100001402, 100001409, 100001412, 100001415, 100001416, 100001417, 100001424, 100001426, 100001432, 100001449, 100001460, 100001464, 100001476, 100001480, 100001483, 100001488;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0462-2026

Related Recalls

Radiography 7300 C; Model Number: 712037;

Philips Medical Systems DMC GmbH

Class I - Dangerous

Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

Oct 1, 2025 Other Medical Devices Nationwide View Details →

Precision CRF; Model Number: 706400;

Philips Medical Systems DMC GmbH

Class I - Dangerous

Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

Oct 1, 2025 Other Medical Devices Nationwide View Details →