IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 System, x-ray, tomography, computed
Class I - DangerousWhat Should You Do?
- Check if you have this product: IntelliSpace Portal DX/HX/EX: 96027,96029,96025,96030,96031,85479,79935, 98087,85175,96024,106013,450003, 950002, 950003,35009,35030,870056, 550012, 85081, 97069, 45001985333, 35036, and 85382. IntelliSpace Portal DX/HX/EX Upgrade: 950100, 55003, and 81178. IntelliSpace Portal DX/HX/EX Demo: 85451 and 87290. IntelliSpace Portal IX: 50306, 40219, and 40368. IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems (Cleveland) Inc
- Reason for Recall:
- Software defect. In certain circumstances, the application may display incorrect measurements of cardiac and aortic anatomy used to select and size the Transcatheter heart valve (THV) to be implanted .
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 System, x-ray, tomography, computed
Product Codes/Lot Numbers:
IntelliSpace Portal DX/HX/EX: 96027,96029,96025,96030,96031,85479,79935, 98087,85175,96024,106013,450003, 950002, 950003,35009,35030,870056, 550012, 85081, 97069, 45001985333, 35036, and 85382. IntelliSpace Portal DX/HX/EX Upgrade: 950100, 55003, and 81178. IntelliSpace Portal DX/HX/EX Demo: 85451 and 87290. IntelliSpace Portal IX: 50306, 40219, and 40368. IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030
Distribution:
Distributed in: US, FL, NC, VT
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0473-2015
Related Recalls
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Philips Medical Systems (Cleveland)
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Philips Medical Systems (Cleveland)
When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.