Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032
Class I - DangerousWhat Should You Do?
- Check if you have this product: S/N :AU2001094-1 UDI: EPTWS0700090 VeriSoft ;EPTWS0700090 BeamAdjust
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- PTW NORTH AMERICA CORPORATION
- Reason for Recall:
- When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032
Product Codes/Lot Numbers:
S/N :AU2001094-1 UDI: EPTWS0700090 VeriSoft ;EPTWS0700090 BeamAdjust
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0479-2022