Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Draeger Medical, Inc.
- Reason for Recall:
- 26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300
Product Codes/Lot Numbers:
UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0501-2022
Related Recalls
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
Draeger Medical