Recall Details

Company:
Draeger Medical, Inc.
Reason for Recall:
26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300

Product Codes/Lot Numbers:

UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0501-2022