Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial nos: 910045-09 911354-02 910045-19 908883-23 910045-15 910045-14 910045-25 908883-04 911354-25 908883-18 908883-11 910045-21 911354-26 908883-06 908883-02 911354-27
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Product Codes/Lot Numbers:
Serial nos: 910045-09 911354-02 910045-19 908883-23 910045-15 910045-14 910045-25 908883-04 911354-25 908883-18 908883-11 910045-21 911354-26 908883-06 908883-02 911354-27
Distribution:
Distributed in: US, AR, CA, KS, KY, MD, MN, NM, OK, PA, RI, TX, UT
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0558-2019
Related Recalls
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS Pheno VE30A and VE40A, Model 10849000
Siemens Medical Solutions USA
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.