Zimmer Spine Ardis Inserter, Rx, non-sterile

Class I - Dangerous
🏥 Medical Devices Recalled: November 12, 2012 Zimmer Other Medical Devices Nationwide

What Should You Do?

  1. Check if you have this product:
    Ardis Inserter, Part number: 3256-01, Lot: 52ZW, 50CL, 53BB, 55AB, 58GS, 58GT, 60JY, 60JZ, 62KP, 63HJ, 64EY, 68AX, 71JE, 75LE Part, 07.00780.701, lot 76CN; part 07.00780.702, lot 76CP;part 07.00900.701, lot 76TR; 07.00900.702, 76TS; 650-0297-01, 70YT; 07.00780.701, 73EC; 07.00780.702, 73EC; 07.00900.701, 76TQ; 650-0239-01, 66PR; 650-0293-01, 70SG ; 650-0293-01, 70SG-R; 650-0179-01, 60CK; 650-0176-01, 60CL; 650-0365-01, 72LP; 07.00780.701, 76CQ; 07.00780.702, 76CR; 07.00780.701, 74EN; 07.00780.702, 74EP; 650-0351-01, 72JH; 650-0351-02, 72JJ; 650-0263-01, 68HL; 650-0208-01, 62FH; 07.00900.701, 76TN; 07.00780.701, 76QG; 07.00780.701, 76LC; 07.00780.702, 76LD;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer, Inc.
Reason for Recall:
Zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation, surgical delays of up to 60 minutes could occur. Sharp edges of the broken implant may contact and damage adjacent structures, requiring surgical intervention. Observed health risks
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Zimmer Spine Ardis Inserter, Rx, non-sterile

Product Codes/Lot Numbers:

Ardis Inserter, Part number: 3256-01, Lot: 52ZW, 50CL, 53BB, 55AB, 58GS, 58GT, 60JY, 60JZ, 62KP, 63HJ, 64EY, 68AX, 71JE, 75LE Part, 07.00780.701, lot 76CN; part 07.00780.702, lot 76CP;part 07.00900.701, lot 76TR; 07.00900.702, 76TS; 650-0297-01, 70YT; 07.00780.701, 73EC; 07.00780.702, 73EC; 07.00900.701, 76TQ; 650-0239-01, 66PR; 650-0293-01, 70SG ; 650-0293-01, 70SG-R; 650-0179-01, 60CK; 650-0176-01, 60CL; 650-0365-01, 72LP; 07.00780.701, 76CQ; 07.00780.702, 76CR; 07.00780.701, 74EN; 07.00780.702, 74EP; 650-0351-01, 72JH; 650-0351-02, 72JJ; 650-0263-01, 68HL; 650-0208-01, 62FH; 07.00900.701, 76TN; 07.00780.701, 76QG; 07.00780.701, 76LC; 07.00780.702, 76LD;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0594-2013

Related Recalls

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

Dec 2, 2025 Other Medical Devices Nationwide View Details →

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.