EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Encore Medical, Lp
- Reason for Recall:
- The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off axis loading which can increase the risk of device failure. This could cause a delay in surgery. If the tip is not seen during surgery, long term risks may include an inflammatory response, infection, and/ or pain
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096
Product Codes/Lot Numbers:
All Lots
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0611-2020
Related Recalls
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
Their is a potential that the reamer may kick or bind up during or immediately prior to use.