Stationary Fluoroscopic X-ray System CombiDiagnost R90, Model 709030 Stationary Fluoroscopic X-ray System

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All CombiDiagnost systems with software version 1.0.0, 1.0.1 and 1.0.2
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Medical Systems Gmbh, DMC
Reason for Recall:
kV/mA lockin not functioning as specified which may result in the patient receiving an increased radiation dose.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Stationary Fluoroscopic X-ray System CombiDiagnost R90, Model 709030 Stationary Fluoroscopic X-ray System

Product Codes/Lot Numbers:

All CombiDiagnost systems with software version 1.0.0, 1.0.1 and 1.0.2

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0612-2019

Related Recalls

CombiDiagnost R90 Software Version R1.0 and R1.1

Philips Medical Systems Gmbh, DMC

Class I - Dangerous

Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.

Jul 22, 2020 Other Medical Devices Nationwide View Details →

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Dec 19, 2019 Diagnostic Equipment Nationwide View Details →

When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.

Dec 19, 2019 Diagnostic Equipment Nationwide View Details →